Quality Control Testing

Quality Control Testing is used to detect errors that
may occur due to test system errors, product defects,
adverse environmental conditions and variance in
operator performance. Ongoing Quality Control Testing
is also used to detect any performance issues of the
system over time. Facility Quality Control Testing Policy
and Procedure should adhere to the manufacturer's
instructions for use and regulatory guidelines.
TRUE METRIX
PRO is a no-coding system, which means
®
the meter does not have to be coded to each lot of test
strips. To assure accurate and reliable results, TRUE
METRIX
PRO offers two kinds of Quality Control Tests.
®
These tests ensure that the TRUE METRIX
is working properly and the user's testing technique
is good.
3 Levels of Control Solution
• Use ONLY TRUE METRIX
perform Control Tests.
• Write date opened on control solution
label. Discard if either 3 months after first
opening or after the date printed next to
the EXP on the bottle label has passed,
whichever comes first.
• Control Test ranges are printed on the
TRUE METRIX
PRO Test Strip vial label.
®
PRO System
®