NIPRO Diagnostics True Metrix Pro Manual page 18

Professional monitoring blood glucose system
Table of Contents

Advertisement

The TRUE METRIX
PRO System was also tested by
®
healthcare professionals at a research center. The data
was compared to parallel results obtained on a Yellow
Springs Instrument (YSI). The table below shows how
often TRUE METRIX
PRO Blood Glucose System venous
®
values obtained by healthcare professionals achieve the
accuracy goals.
Venous samples drawn into sodium heparin anticoagu-
lant tubes.
Venous Blood < 75 mg/dL
Within ±5 mg/dL
5/6 = 83.3%
Within ±10 mg/dL 6/6 = 100%
Within ±15 mg/dL 6/6 = 100%
Venous Blood ≥ 75 mg/dL
Within ±5%
20/106 = 18.9%
Within ±10%
60/106 = 56.6%
Within ±15%
101/106 = 95.3%
Within ± 20%
106/106 = 100%
TRUE METRIX
PRO System Comprehensive Resource Guide (GDH-FAD Enzyme)
®
Performance Characteristics
for Healthcare Professionals
Precision (Repeatability):
Precision describes the variation between results.
Precision results were performed in a laboratory.
Blood (Within Lot): N=100
Mean (mg/dL)
SD (mg/dL)
CV%
Control Solution: N=100
Mean (mg/dL)
SD (mg/dL)
CV%
44
89
150
199
1.8
3.6
4.9
7.3
4.1
4.0
3.3
3.7
33
105
306
1.6
3.8
10.3
4.8
3.6
3.4
Performance Characteristics 13
4
cont.
329
9.6
2.9