Probe Head Replacement 
The probe head (on the tip of the probe) holds a thin Teflon® membrane and an electrolyte 
container. Calibration errors or reading errors could indicate loss of electrolyte or a 
dirty/damaged membrane. The membrane is very delicate and can easily be damaged if it 
comes in contact with solid objects. Replacement probe heads are available from an Extech 
distributor.  
Replacement procedure 
1. Unscrew the probe head and empty the old electrolyte from the container. 
2. Fill the new probe head/electrolyte container with electrolyte. 
3. Screw the probe head onto the probe handle. Excess electrolyte may be forced out through 
the threads. Wipe clean. 
4. Lightly tap the probe to remove any air bubbles. 
 