Display Indications - Leica IP S Instructions For Use Manual

Printer for slides
Table of Contents

Advertisement

5.2

Display indications

Display indication
Display
Leica IP S
Magazine empty (in combination with MAG. EMPTY LED)
Magazine No. 1 empty
1 -
2 -
Magazine No. 2 empty
3 -
Magazine No. 3 empty
If
MANUAL FEED
has been selected in the printer driver settings,
appear in the display after the print job has been sent. The printer will
wait for an individual specimen slide to be placed in the feeding chute
for printing.
Status messages
00
Ink print head cleaning in progress.
11
Temperature in the flash power supply is too high.
Printer is too hot and initiates a short cool-down period.
The print job will be resumed automatically after a short
period of time. To prevent frequent job interruptions due to
heat build-up, make sure the ventilation grids of the printer
are unobstructed and keep the printer away from other heat
sources.
Consider operating the printer in an air-conditioned room. If the
temperature does not drop to a value within the allowed range
within 10 minutes,
let it cool; check ambient temperature.
13
Flash bulb has reached its maximum life.
The flash bulb has reached the end of its specified service life
and must be replaced.
If this message is ignored, the resistance of subsequent
printouts can be affected.
14
Prompt requesting maintenance.
If this message is displayed, the instrument will be due for
maintenance within the next few weeks. Confirm the prompt by
pressing ERROR.
After about 3 weeks the message will be displayed again and
will not disappear from the screen when pressing ERROR.
You can still continue printing but maintenance
necessary.
55
is displayed. Switch the instrument off and
5
Operation
0
will
is absolutely
49

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents