H43.0.01.6C-11 Operating Manual GOX 20 Page 3 of 6 2.3 Safety guidelines This device has been designed and tested in accordance with the safety regulations for electronic devices. However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using the device.
11,6 11,3 If the value displayed on the GOX 20 differs from the one given in the table, it may be adjusted by means of the „slope-knob“. If it can no longer be adapted to the values given, the oxygen electrode needs maintenance or is used up and needs to be replaced.
H43.0.01.6C-11 Operating Manual GOX 20 Page 5 of 6 The oxygen electrode The oxygen electrode is an active electrode consisting of a silver cathode and a lead anode using potassium hydroxide (KOH) as an electrolyte. In case of oxygen being present it will be reduced at the silver cathode and the electrode produces electric current.
H43.0.01.6C-11 Operating Manual GOX 20 Page 6 of 6 Specification Measuring range Oxygen 0.0 ... 20.0 mg/l O Temperature 0.0 ... 40.0 °C Resolution Oxygen 0.1 mg/l O Temperature 0.1 °C Accuracy ( at nominal temperature) ±1digit) Oxygen ± 2 % of meas. value ± 0.2 mg/l Temperature ±...