H40.0.01.6C-09 Operating Manual GMH 3330 Page 3 of 16 General Note Read this document carefully and get used to the operation of the device before you use it. Keep this document within easy reach near the device for consulting in case of doubt.
_DANGER 3 Product Specification 3.1 Scope of supply The scope of supply includes: Device GMH 3330, incl. 9V battery block Operation manual 3.2 Operation and maintenance advice • Battery operation: If ‘bAt’ is shown in the lower display the battery has been used up and needs to be replaced. However, the device will operate correctly for a certain time.
H40.0.01.6C-09 Operating Manual GMH 3330 Page 5 of 16 Treat device and sensor carefully. Use only in accordance with above specification. (do not throw, hit against etc.). Protect plug and socket from soiling. To disconnect thermocouple sensor plug do not pull at the cable but at the plug.
H40.0.01.6C-09 Operating Manual GMH 3330 Page 6 of 16 Messages at device startup: The device will show message at the startup depended on the connected sensor. 4.2 Pushbuttons On/off key min/max when taking measurements: press shortly: min. or max. measuring value will be displayed press for 1 sec.: the min.
H40.0.01.6C-09 Operating Manual GMH 3330 Page 7 of 16 4.4 Pop-up clip Handling: • Pull at label “open” in order to swing open the pop-up clip. • Pull at label “open” again to swing open the pop-up clip further. Pop-up clip at position 90°...
Page 8
H40.0.01.6C-09 Operating Manual GMH 3330 Page 8 of 16 ’AVG’: Selection of Averaging Proceedings for Flow L o g g r.H. Measurement (only STS005/020) continuous averaging - the average value calculated from Cont: the measurements conducted during the averaging period will be displayed.
H40.0.01.6C-09 Operating Manual GMH 3330 Page 9 of 16 Measurements Using Combination Measuring Sensor TFS0100 The TFS0100 has been especially designed to carry out measurements of ambient temperature. All TFS0100-probes are interchangeable without recalibration being required. The scope of supply includes one sensor to measure relative atmospheric humidity and another one to measure the ambient tempera- ture T1.
H40.0.01.6C-09 Operating Manual GMH 3330 Page 10 of 16 Measurements Using Flow Measuring Probes STS005 / STS020 Two types of measuring probes are available for flow speed measurements: Please note: -use STS 005 to measure water flow -use STS 020 to measure air flow Incorrect use will result in incorrect measurements! Please observe max.
H40.0.01.6C-09 Operating Manual GMH 3330 Page 11 of 16 Notes to Special Functions 8.1 Zero Displacement (’Offset’) A zero displacement can be carried out for each of the two temperature channels T1 (TFS0100 only) and T2: temperature displayed = temperature measured - offset Standard setting: 'off' = 0.0°, i.e.
H40.0.01.6C-09 Operating Manual GMH 3330 Page 12 of 16 How to carry out calibration Please note: the calibration is only possible, if the logger memory is empty. Start calibration: press "CAL" (key 3) for 2 sec. (after more than 10 sec. the factory calibration will be set) The display prompts you to measure the first humidity value.
H40.0.01.6C-09 Operating Manual GMH 3330 Page 14 of 16 11 The serial interface By means of the serial interface and a suitable electrically isolated interface adapter (USB 3100, USB 3100 N, GRS 3100 or GRS 3105) the device can be connected to a computer for data transfer.
H40.0.01.6C-09 Operating Manual GMH 3330 Page 16 of 16 13 Reshipment and Disposal 13.1 Reshipment All devices returned to the manufacturer have to be free of any residual of measuring media and other hazardous substances. Measuring residuals at housing or sensor may be a risk for...