H70.0.21.6C-01 Operating Manual GMH 3511 page 3 of 14 1 General Note Read this document carefully and get used to the operation of the device before you use it. Keep this document within easy reach near the device for consulting in case of doubt.
H70.0.21.6C-01 Operating Manual GMH 3511 page 4 of 14 If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked accordingly to avoid re-starting. Operator safety may be a risk if:...
H70.0.21.6C-01 Operating Manual GMH 3511 page 5 of 14 4 Handling 4.1 Display Main display: pH value, ORP value (mV, mV Display elements to show minimum/maximum/ memorized measuring value Arrows to selected measuring unit mbar Warning signal (low battery or recalibration prompt)
H70.0.21.6C-01 Operating Manual GMH 3511 page 7 of 14 5 Start Operation Connect electrodes, turn device on via key. After segment test the device is ready for measuring. Remove protective cap from electrode. (Attention: Cap should contain KCL 3 M or storage solution) 6 Principles of the measurements 6.1 pH measurement...
H70.0.21.6C-01 Operating Manual GMH 3511 page 8 of 14 6.3 pH electrode 6.3.1 Design In most cases so-called combination electrodes are used. That means that all needed elements are integrated in a single electrode (including reference electrode). Sometimes even a temperature sensor is integrated.
H70.0.21.6C-01 Operating Manual GMH 3511 page 9 of 14 6.4 Calibration of pH measurement The electrode data of pH electrodes are subject to fluctuation due to ageing and manufacturing tolerances. Therefore it is necessary to check the calibration with buffer solutions before measurements take place. If deviations are too large, a recalibration is necessary.
H70.0.21.6C-01 Operating Manual GMH 3511 page 10 of 14 Error messages of pH calibration: Neutral buffer not permissible Always use neutral buffer as first solution - Wrong buffer solution (exception: 1 point calibration) Use new buffer solution - Buffer solution defective...
H70.0.21.6C-01 Operating Manual GMH 3511 page 11 of 14 8 Output The output can be used as serial interface (for USB 3100, USB 3100 N, GRS 3100 or GRS 3105 interface adapters) or as analog output (0-1V). If none of both is needed, we suggest to switch the output off, because battery life then is extended.
H70.0.21.6C-01 Operating Manual GMH 3511 page 12 of 14 10 Error and System Messages Display Description What to do? Battery empty Replace battery No display or Mains operation: wrong voltage or polarity Check power supply, replace it when necessary confused...
H70.0.21.6C-01 Operating Manual GMH 3511 page 14 of 14 13 Notes A: temperature influence on pH buffer solutions GPH buffer capsules for 100 ml buffer solution Capsules for do-it-yourself mixing – unopened capsules can be stored a long time (approx. 3 years) T [°C]...