Summary of Contents for Vanguard Instruments CT-7000 S3
Page 1
CT-7000 S3 DIGITAL CIRCUIT BREAKER ANALYZER USER’S MANUAL Vanguard Instruments Company, Inc. 1520 S. Hellman Ave. Ontario, California 91761, USA TEL: (909) 923-9390 August 2013 FAX: (909) 923-9391 Revision 1...
Page 2
SAFETY WARNINGS AND CAUTIONS Only trained operators shall use the CT-7000 S3. All devices under test shall be off-line and fully isolated. Do not perform test procedures or service unless another person is also present who is capable of rendering aid and resuscitation.
3.4.1. Saving a Timing Record in Flash EEPROM .............. 77 3.4.2. Printing a Directory of Test Records Stored in the CT-7000 S3's Memory .... 78 3.4.3. Printing a Directory of Test Records Stored in a USB Flash Drive ......80 3.4.4.
Page 4
3.5.1. Recalling a Breaker Test Plan for Use ..............93 3.5.2. Printing a Directory of Test Plans Stored in the CT-7000 S3’s Memory ....96 3.5.3. Printing a Directory of Test Plans Stored in a USB Flash Drive ......98 3.5.4.
Page 5
REV 1 CT-7000 S3 USER’S MANUAL LIST OF FIGURES Figure 1. CT-7000 S3 Controls and Indicators ................. 6 Figure 2. Typical 3-Phase Circuit Breaker Connections ..............10 Figure 3. Typical Connections for Series Contact Circuit Breaker ..........11 Figure 4. Typical DC Trip and DC Close Control Circuit Connection ..........12 Figure 5.
CONVENTIONS USED IN THIS DOCUMENT This document uses the following conventions: [KEY] [SWITCH] • A key or switch on the CT-7000 S3 is indicated as • Menu options are referenced as (MENU OPTION). • Screen and menu names are referenced as “SCREEN/MENU NAME”.
1 second, 10 seconds, and 20 seconds. Contact Timing Input Channels The CT-7000 S3's dry contact timing channels (up to 6 channels) are used to time the circuit breaker main contacts. Each main contact timing channel is capable of detecting the main contact and insertion resistor contact time.
Pass/Fail report is generated by comparing the actual performance with the specifications in the stored test plan. Test plans can be transferred to the CT-7000 S3 from a USB Flash drive, or from a PC via the USB port or Bluetooth (optional).
Dual Ground Testing Mode With the optional dual ground testing mode, the CT-7000 S3 can measure a CB contact time with ground being applied to both sides of the bushings. The clamp-on probe is connected to one side of the CB safety ground straps, and an AC signal is coupled to this strap.
CABLES furnished with full set of test leads OPTIONS transportation cases (available for the CT-7000 S3 and travel transducers), on-line timing mode, dual ground testing mode, bluetooth interface SAFETY designed to meet UL 6101A-1 and CAN/CSA C22.2 No 1010.1-92 standards ENVIRONMENT Operating: -10°C to +50°C (+15°F to +122°F)
CT-7000 S3 USER’S MANUAL CT-7000 S3 Controls and Indicators The CT-7000 S3’s controls and indicators are shown in Figure 1 below. A leader line with an index number points to each control and indicator, which is cross-referenced to a functional description in Table 2.
CT-7000 S3 USER’S MANUAL REV 1 Table 2. Functional Descriptions of CT-7000 S3 Controls and Indicators Item Panel Markings Functional Description Number CONTACT INPUT Female connectors for the contact channels. The CT-7000-6 S3 is pictured (C1-C6) with 6 channels. T1, T2, T3 Digital travel transducer channels (16-pin connectors).
Printer Paper The CT-7000 S3’s built-in thermal printer uses 4.5-inch wide thermal paper for printing test results. To maintain the highest print quality and to avoid paper jams, the use of thermal paper supplied by Vanguard Instruments Company is highly recommended. Additional paper can be ordered from the following sources: Vanguard Instruments Co, Inc.
Page 14
CT-7000 S3 USER’S MANUAL REV 1 Replacing the Thermal Printer Paper The roll of thermal paper is housed inside a dispenser underneath the printer cover. To replace the paper, follow the steps below: • Unscrew the two large printer cover screws and remove the printer cover.
REV 1 CT-7000 S3 USER’S MANUAL OPERATING PROCEDURES Cable Connections 3.1.1. Contact Cable Connections A typical contact cable connection to a circuit breaker is shown in Figure 2. Red and black clips are connected across the circuit contact phases A, B, and C. A typical circuit breaker with series contacts is shown in Figure 3.
CT-7000 S3 USER’S MANUAL 3.1.2. Initiate Cable Connections The CT-7000 S3 can trip or close breakers through a solid-state device operating on any AC or DC control voltage ranging from 10 to 300 Volts. Both the trip and close circuits are protected by 5 Ampere fuses.
The CT-7000 S3 On-Line timing mode requires the use of 3 AC current sensors (provided with the Online Timing Option, Part no. 9021-OT) to monitor the contact time phases A, B, and C.
3.1.4. Dual Ground Connections (with optional Dual Ground feature) With the optional dual ground testing mode, the CT-7000 S3 can measure a CB contact time with ground being applied to both sides of the bushings. The clamp-on probe is connected to one side of the CB safety ground straps, and an AC signal is coupled to this strap.
CT-7000 S3 USER’S MANUAL REV 1 3.1.5. Analog and Digital Voltage Monitoring Connections The analog voltage input “V1” can monitor a breaker’s DC control voltage during an operation. The analog voltage input records the nominal DC voltage at no load and the minimum DC voltage while the Trip or Close coil is energized.
3.1.6. External Trigger Input Connections The External Trigger Mode can be used to start recording data when the CT-7000 S3 senses a voltage. A typical application for the External Trigger Mode is to time a circuit breaker in a Close operation and to start timing only when the Close coil is energized, thus bypassing the 52X relay delay time.
CT-7000 S3 USER’S MANUAL REV 1 3.1.7. Digital Transducer Connection A typical digital transducer connection is shown in Figure 11. Figure 11. Typical Digital Transducer Connection...
The CT-7000 S3 provides three channels for resistor type transducers. The transducer resistance input can range from 200 Ohms to 10K Ohms. The CT-7000 S3 supplies a 5 Vdc reference voltage to power the resistor type transducer. The sense voltage is translated into a travel distance based on the transducer calibration parameters.
CT-7000 S3 USER’S MANUAL REV 1 Changing Setup Parameters 3.2.1. Entering Test Record Header Information You can enter the test record header information before performing tests. The record header includes identifying information such as the company, station, circuit, model number, etc. Once the header information has been entered, it will apply to all subsequent test records.
Page 27
REV 1 CT-7000 S3 USER’S MANUAL d. The following screen will be displayed: STATION: ↑↓ to position "ENTER" TO ACCEPT [ENTER] Type the station name using the keypad and then press the key. e. The following screen will be displayed: CIRCUIT: ↑↓...
Page 28
CT-7000 S3 USER’S MANUAL REV 1 h. The following screen will be displayed: SERIAL NUMBER: ↑↓ to position "ENTER" TO ACCEPT [ENTER] Type the circuit breaker serial number using the keypad and then press the key. i. The following screen will be displayed: OPERATOR: ↑↓...
REV 1 CT-7000 S3 USER’S MANUAL 3.2.2. Setting the Clock Follow the steps below to set the CT-7000 S3’s internal clock: a. Start from the “START-UP” menu: 1. TIME BREAKER 09/12/13 2. GET RESULTS 10:10:10 3. SAVE/RESTORE 4. SETUP 5. DIAGNOSTICS Press the key (SETUP).
REV 1 3.2.3. Configuring Automatic or Manual Printing of Test Results The CT-7000 S3 can be configured to print graphs and tabulated results automatically or manually after each test. To configure the printing method: a. Start from the “START-UP” menu: 1.
CT-7000 S3 USER’S MANUAL 3.2.4. Setting the Units of Measure The CT-7000 S3 supports and displays both English and Metric calculations. You can switch between the English and Metric systems using the steps below: a. Start from the “START-UP” menu: 1.
CT-7000 S3 USER’S MANUAL REV 1 3.2.5. Setting the OPEN Timing Analysis Points Two analysis points are used to calculate the velocity of the circuit breaker in the OPEN operation. To configure the OPEN timing analysis points: a. Start from the “START-UP” menu: 1.
Page 33
REV 1 CT-7000 S3 USER’S MANUAL Percentage of stroke is the distance based upon the percentage of the total breaker’s stroke distance. The distance is always measured from the starting point at the fully closed position of the breaker contacts.
Page 34
CT-7000 S3 USER’S MANUAL REV 1 e. The following screen will be displayed: OPEN ANALYSIS PT 2 1. PERCENT OF STROKE 2. DIST FROM CLOSE 3. CONTACT +/- TIME 1. PERCENT OF STROKE Press the key (PERCENT OF STROKE) to set the open analysis point #2 as a percentage of the total stroke value.
• For complex velocity calculations, the user can create a test plan using the included CT-7000 S3 Breaker Analysis PC Software. The test plan can then be downloaded to the CT-7000 S3 and recalled before running timing tests.
Page 36
CT-7000 S3 USER’S MANUAL REV 1 The contact time is printed on the tabulated test results printout in both milliseconds and cycles. The cycle readings can be in 50 Hz or 60 Hz. You can select the preferred frequency using the steps below: a.
Page 37
60 hz set Press any key to return to the “START-UP” menu. Sample tabulated test results are shown in Figure 13 and Figure 14. If the frequency value is changed, the cycle values are automatically re- calculated by the CT-7000 S3. NOTE...
CT-7000 S3 USER’S MANUAL REV 1 Frequency Set to 50 Hz Cycle readings at 50 Hz Figure 13. Typical 50 Hz Tabulated Test Results Frequency set to 60 Hz Cycle readings at 60 Hz Figure 14. Typical 60 Hz Tabulated Test Results...
CT-7000 S3 USER’S MANUAL 3.2.7. Configuring the Channel Settings The CT-7000 S3 is available with 3 or 6 contact inputs. Since most common timing applications require the use of only 3 contact timing channels and one travel transducer channel, there is no need to print data for more than 3 timing channels and one transducer channel on the graphic and tabulated reports.
Page 40
CT-7000 S3 USER’S MANUAL REV 1 Press either the key (3 CONTACTS) or the key (6 CONTACTS) to select the corresponding number of contact channels. d. The following screen will be displayed: 1. transducer 1 2. transducers 1,2 3. transducers 1,2,3...
The value can be between 1 and 300. A filter setting of 1 allows the CT-7000 S3 to pick up the first contact transition time after the resistor contact activity is detected. A filter setting of 300 allows the CT-7000 S3 to pick up the last contact transition time after the resistor contact activity is detected.
CT-7000 S3 USER’S MANUAL REV 1 Filter Setting @ 1 will select this contact time Filter Setting @ 300 will select this contact time Filter Setting @ 1 will select this contact time Filter Setting @ 300 will select this contact time...
CT-7000 S3 USER’S MANUAL 3.2.9. Configuring the Transducer Encoder Filter Setting In a typical 1-second timing record, the CT-7000 S3 records 20,000 data points for each of the contact channels, digital transducer channels, voltage input channels, CT channel, DCR channel, resistor transducer channel, and initiate current channel.
Page 44
The following screen will be displayed: ENTER FILTER TIME (010 - 999) [ENTER] Type the filter time (in milliseconds) using the keypad, and then press the to return to the “START-UP” menu. The CT-7000 S3 will not record any transducer encoder data after this time.
REV 1 CT-7000 S3 USER’S MANUAL 3.2.10. Configuring the Digital Rotary Transducer Settings A rotary transducer requires the user to enter the defined linear distance in millimeters or inches per one degree of rotary motion. Follow the steps below to configure the settings for a digital rotary transducer: a.
Page 46
CT-7000 S3 USER’S MANUAL REV 1 1. ENGLISH (In./deg) Press the key to enter the rotary encoder linear distance per degree using English units (inches per degree). The following screen will be displayed: inches/degree 1.442 in/deg [CLEAR] Type the desired value using the keypad. You can press the...
The CT-7000 S3 provides three resistor type transducer channels. The transducer must be configured before it can be used with the CT-7000 S3. Up to 9 resistor transducer setups can be stored in the CT-7000 S3’s Flash EEPROM. When a resistive transducer is used with the CT-7000 S3, it is shown as transducer #1 on the timing report as shown in Figure 17.
Page 48
CT-7000 S3 USER’S MANUAL REV 1 d. The following screen will be displayed: RESISTIVE ENCODER: 1. CREATE NEW SETUP 2. LOAD SETUP 3. PRINT SETUP DIR Press the key (CREATE NEW SETUP). e. The following screen will be displayed: RESISTIVE ENCODER: 1.
Page 49
REV 1 CT-7000 S3 USER’S MANUAL h. The following screen will be displayed (the units of measure displayed will depend on your choice in step e): ENTER THE DISTANCE MOVED: "ENTER TO CONTINUE" Enter the distance the transducer was moved using the keypad, and then press the [ENTER] key.
Page 50
CT-7000 S3 USER’S MANUAL REV 1 l. The setup will be saved and the following confirmation screen will be displayed: SETUP SAVED! Press any key to return to the “START-UP” menu. Loading a Resistor Type Transducer Setup Follow the steps below to load a resistor type transducer setup: a.
Page 51
REV 1 CT-7000 S3 USER’S MANUAL d. The following screen will be displayed: RESISTIVE ENCODER: 1. CREATE NEW SETUP 2. LOAD SETUP 3. PRINT SETUP DIR Press the key (LOAD SETUP). e. The following screen will be displayed: ENTER SETUP NUMBER TO LOAD (1-9) Type the setup number to load using the numeric keypad.
Page 52
CT-7000 S3 USER’S MANUAL REV 1 Printing a Resistor Type Transducer Setup Directory You can print a directory of the stored resistor type transducer setups on the built-in thermal printer using the steps below: a. Start from the “START-UP” menu: 1.
REV 1 CT-7000 S3 USER’S MANUAL Performing Circuit Breaker Timing Tests The CT-7000 S3 can initiate the breaker operation and perform a timing test on the following operations: • OPEN • CLOSE • OPEN-CLOSE • CLOSE-OPEN • OPEN-CLOSE-OPEN The CT-7000 S3 can start the OPEN-CLOSE operations without a delay or by using a programmable delay between the OPEN and CLOSE commands.
REV 1 3.3.1. Timing an OPEN Operation The CT-7000 S3 can time breakers with or without insertion resistors. The insertion resistance can range from 10 to 5,000 Ohms. Any insertion resistance greater than 5,000 Ohms is detected as an open circuit. The timing results will show the main contact time and the insertion resistor contact time.
Page 57
REV 1 CT-7000 S3 USER’S MANUAL resistor value: 1. LESS THAN 1000 OHMS 2. 1000 to 2000 OHMS 3. MORE THAN 2000 OHMS Select the resistance value by pressing the corresponding key ( , or Continue to step d. d. The following screen will be displayed: TIMING WINDOW: 1.
Page 58
“START-UP” menu once testing is finished. • Please see section 3.3.5 for information about printing test results. • The CT-7000 S3 can be configured to print tabulated and graphics reports automatically after performing a test. Please see section 3.2.3 for further NOTES information.
There are three options when timing a CLOSE-OPEN operation: 1. Contact #1 CLOSE The CT-7000 S3 will initiate a CLOSE operation. The open operation is then initiated after contact channel #1 is closed. This option closes simulates the breaker closing on a fault condition in the field.
Page 60
CT-7000 S3 USER’S MANUAL REV 1 c. The following screen will be displayed: INSERTION RESISTOR? 1. NO 2. YES Press the key (NO). d. The following screen will be displayed: TIMING WINDOW: 1. WINDOW = 1 2. WINDOW = 10 SEC 3.
Page 61
REV 1 CT-7000 S3 USER’S MANUAL g. The following screen will be displayed: C-O SECOND TRIGGER: 1. CONTACT #1 CLOSE 2. SET DELAY 3. NO DELAY Contact #1 CLOSE Press the key if you would like the open operation to be initiated after contact channel #1 is closed.
Page 62
CT-7000 S3 USER’S MANUAL REV 1 i. The following screen will be displayed: TEST IN PROGRESS HOLD "ARM" UNTIL TEST COMPLETES. (UP TO 25 SECONDS) [ARM] Continue to hold down the switch until testing is finished. You will be returned...
REV 1 CT-7000 S3 USER’S MANUAL 3.3.3. Timing an OPEN-CLOSE-OPEN Operation The OPEN-CLOSE-OPEN operation requires the user to enter two time delays (in milliseconds) between the circuit breaker operations. The first delay is from the first OPEN command to the CLOSE command, and the second delay is from the CLOSE command to the second OPEN command.
Page 64
CT-7000 S3 USER’S MANUAL REV 1 e. The following screen will be displayed: TRIGGER MODE: 1. INTERNAL TRIGGER 2. EXTERNAL TRIGGER Press the key (INTERNAL TRIGGER). f. The following screen will be displayed: TIMING MODE: 1. OPEN 2. CLOSE 3. O-C 4.
Page 65
REV 1 CT-7000 S3 USER’S MANUAL i. The following screen will be displayed: HOLD "ARM" UNTIL TEsT COMPLETES. "START" TO BEGIN "STOP" TO ABORT [ARM] [START] Hold down the switch and press the key. j. The following screen will be displayed: TEST IN PROGRESS HOLD "ARM"...
Page 66
CT-7000 S3 USER’S MANUAL REV 1 3.3.4. Performing an On-Line Timing Test on Live Contacts The optional "On-line" timing mode (part no. 9021-OT) can be used to capture the breaker's trip or close time, the trip/close coil current “fingerprint”, and the battery supply voltage while the breaker is still in service.
Page 67
REV 1 CT-7000 S3 USER’S MANUAL Press the key (OPEN). e. The following screen will be displayed: -> NO INITIATE CABLES -> EXTERNAL TRIGGER ONLY! "START" TO BEGIN "STOP" TO ABORT [START] Press the key. f. The following screen will be displayed temporarily: TESTING IN PROGRESS...
CT-7000 S3 USER’S MANUAL REV 1 3.3.5. Printing or Viewing Timing Results Follow the steps below to print or view the timing results after performing a circuit breaker timing test: a. After performing a timing test you will be returned to the “START-UP” menu: 1.
Page 73
REV 1 CT-7000 S3 USER’S MANUAL Press the key (DISPLAY RESULTS) to display the test results on the unit’s LCD screen. The following screen will be displayed: UP / DWN ARROWS TO SCROLL RESULTS... "STOP" TO EXIT... [] [] Press the keys to scroll through the test results.
CT-7000 S3 USER’S MANUAL REV 1 Enter the end point for the expansion by typing the first digit of the 100 millisecond increment. For example, to select 900ms, press the key. The screen will be updated as shown: EXPANSION FROM: 900MS "ENTER"...
REV 1 CT-7000 S3 USER’S MANUAL Table 3. Descriptions of Tabulated Test Results Elements Item Description Number Test record number and the date and time the test was performed. Test record header information (Company, Station, Circuit, etc.). Type of test performed (OPEN, CLOSE, O-C, C-O, or O-C-O).
CT-7000 S3 USER’S MANUAL REV 1 Figure 27. Graphical Interpretation of an OPEN Timing Shot Figure 28. Graphical Interpretation of a CLOSE Timing Shot...
REV 1 Working with Test Records 3.4.1. Saving a Timing Record in Flash EEPROM After performing a test, you can store the results in the CT-7000 S3’s Flash EEPROM. To save a test record: a. Start from the “START-UP” menu: 1.
CT-7000 S3 USER’S MANUAL 3.4.2. Printing a Directory of Test Records Stored in the CT-7000 S3's Memory You can print a directory of all the test records stored in the CT-7000 S3’s Flash EEPROM by using the steps below: a. Start from the “START-UP” menu: 1.
REV 1 CT-7000 S3 USER’S MANUAL 3.4.3. Printing a Directory of Test Records Stored in a USB Flash Drive To print a directory of all test records stored in a USB Flash drive: a. Make sure the USB Flash drive is inserted in the unit's USB Flash drive port ("USB MEM"...
CT-7000 S3 USER’S MANUAL REV 1 3.4.4. Recalling a Test Record from the Flash EEPROM Follow the steps below to recall a test record from the CT-7000 S3’s Flash EEPROM: a. Start from the “START-UP” menu: 1. TIME BREAKER 09/17/13 2.
Page 87
REV 1 CT-7000 S3 USER’S MANUAL e. The following screen will be displayed: RESTORE COMPLETE PRINT RECORD? 1. YES 2. NO Press the key (YES) if you would like to print the restored test record. The following screen will be displayed: 1.
REV 1 3.4.5. Restoring a Test Record from a USB Flash Drive You can restore a test record from a USB Flash drive to the CT-7000 S3's working memory using the steps below: a. Make sure the USB Flash drive containing the test record(s) is inserted in the CT-7000 S3's USB Flash drive port ("USB MEM"...
Page 89
REV 1 CT-7000 S3 USER’S MANUAL [ENTER] Type the record number that you would like to restore and press the key. If you do not know the record number, you can print a test record directory using the instructions in section 3.4.3.
REV 1 3.4.6. Copying Test Records to a USB Flash Drive You can easily copy test records stored in the CT-7000 S3's Flash EEPROM to a connected USB Flash drive using the steps below: a. Make sure a USB Flash drive is inserted in the unit's Flash drive port ("USB MEM" port).
Page 91
REV 1 CT-7000 S3 USER’S MANUAL Type the record number that you would like to copy to the USB Flash drive and then [ENTER] press the key. If you do not know the record number, you can first print a test record directory using the instructions in section 3.4.2.
CT-7000 S3 USER’S MANUAL REV 1 3.4.7. Deleting Test Records from the Flash EEPROM You can delete one or all test records stored in the CT-7000 S3’s Flash EEPROM. To delete test records: a. Start from the “START-UP” menu: 1. TIME BREAKER 09/17/13 2.
Page 93
ERASE COMPLETE Press any key to return to the “START-UP” menu. ERASE ALL SHOTS! Press the key to erase all test records from the CT-7000 S3’s Flash EEPROM. The following screen will be displayed: ALL SHOTS WILL BE ERASED! "ENTER" TO CONTINUE...
Page 94
CT-7000 S3 USER’S MANUAL REV 1 The following screen will be displayed when all the test records have been completely erased: ERASE COMPLETE ANY KEY TO CONTINUE Press any key to return to the “START-UP” menu.
REV 1 CT-7000 S3 USER’S MANUAL 3.4.8. Deleting Test Records from a USB Flash Drive To erase one or all test records stored on a USB Flash drive: a. Make sure a Flash drive is inserted in the unit's Flash drive port ("USB MEM" port). Then start from the "START-UP"...
Page 96
CT-7000 S3 USER’S MANUAL REV 1 ERASE THUMB DRIVE REC_ [ENTER] Type the record number to be deleted and press the key. If you do not know the record number, you can first print a test record directory using the instructions in section 3.4.3.
Page 97
REV 1 CT-7000 S3 USER’S MANUAL ALL THUMB DRIVE RECORDS ERASED! Press any key to return to the "START-UP" menu.
The CT-7000 S3 comes with the Vanguard Circuit Breaker Analysis S2 (VCBA S2) PC Software that can be used to create circuit breaker test plans. The test plans can then be transferred to the CT-7000 S3. The CT-7000 S3 can store up to 99 circuit breaker test plans in its Flash EEPROM.
Page 99
3. PRINT PLAN DIR 4. ERASE TEST PLAN 5. SAVE TEST PLAN Press the key (LOAD TEST PLAN). e. The first test plan stored in the CT-7000 S3’s Flash EEPROM will be listed: TEST PLAN # 01 BZO-145-20-7 SIEMENS DEMO TEST PLAN 1 []...
CT-7000 S3 USER’S MANUAL REV 1 NOTE NO TEST PLANS FOUND "ENTER" TO CONINUE [ENTER] Press the key to return to the “START-UP” menu. Pass/Fail Indicators Figure 31. Typical Test Results Printout with Pass/Fail Indicators...
CT-7000 S3 USER’S MANUAL 3.5.2. Printing a Directory of Test Plans Stored in the CT-7000 S3’s Memory You can print a directory of all the test plans stored in the CT-7000 S3’s Flash EEPROM using the steps below: a. Start from the “START-UP” menu: 1.
REV 1 CT-7000 S3 USER’S MANUAL 3.5.3. Printing a Directory of Test Plans Stored in a USB Flash Drive Follow the steps below to print a directory of all test plans stored in a USB Flash drive: a. Make sure a USB Flash drive is inserted in the unit's USB Flash drive port ("USB MEM"...
Page 104
CT-7000 S3 USER’S MANUAL REV 1 e. The following screen will be displayed: 1. INTERNAL DIRECTORY 2. THUMB DRIVE DIR Press the key (THUMB DRIVE DIR). The directory of test plans stored on the USB Flash drive will be printed on the built-in thermal printer and you will be returned to the...
CT-7000 S3 USER’S MANUAL 3.5.4. Printing a Breaker Test Plan Stored in the CT-7000's Memory Follow the steps below to print a breaker test plan stored in the CT-7000 S3’s Flash EEPROM: a. Start from the “START-UP” menu: 1. TIME BREAKER 09/18/13 2.
Page 106
CT-7000 S3 USER’S MANUAL REV 1 e. The first test plan stored in the CT-7000 S3’s Flash EEPROM will be listed: TEST PLAN # 01 BZO-145-20-7 SIEMENS DEMO TEST PLAN 1 [] [] You can scroll through the list of test plans by pressing the keys.
REV 1 CT-7000 S3 USER’S MANUAL 3.5.5. Printing a Breaker Test Plan Stored in a USB Flash Drive Follow the steps below to print a breaker test plan stored in a USB Flash drive: a. Make sure a USB Flash drive is connected to the unit's USB Flash drive port ("USB MEM"...
Page 108
CT-7000 S3 USER’S MANUAL REV 1 1. PRINT INTERNAL TP 2. PRINT FLASH DRV TP Press the key (PRINT FLASH DRV TP). f. The following screen will be displayed: PRINT THUMB DRIVE TP PLAN_ [ENTER] Type the test plan number to be deleted and press the key.
REV 1 CT-7000 S3 USER’S MANUAL 3.5.6. Copying a Test Plan to a USB Flash drive Follow the steps below to copy a test plan from the unit's internal memory to a connected USB Flash drive: a. Make sure a USB Flash drive is inserted in the unit's USB Flash drive port ("USB MEM"...
Page 110
CT-7000 S3 USER’S MANUAL REV 1 ENTER TP NUMBER TO COPY TO FLASH DRV TP NUMBER: [ENTER] Type the test plan number to copy using the keypad and then press the key. If you do not know the test plan number, you can first print a test plan directory using the instructions in section 3.5.2.
DIAGNOSTICS, VERIFICATION, AND TROUBLESHOOTING Performing a Slow-Close Test The CT-7000 S3 offers a unique “Slow-Close” test feature. This test measures the distance a breaker’s contact travels from the fully opened position to the point of contact or “touch” position, and the contact wipe or penetration distance. The test requires the operator to manually jack the breaker’s contact from the fully open position to the fully closed position.
REV 1 CT-7000 S3 USER’S MANUAL d. The following screen will be displayed: CONTACT 1: OPEN CONTACT 2: OPEN CONTACT 3: OPEN LIFT TO CLS BRKR NOW Manually start closing the breaker and then press the "ENTER" key. The test results will be printed and you be returned to the “START-UP”...
CT-7000 S3 USER’S MANUAL REV 1 Performing a Transducer Self-Test You can check the transducer electronics by connecting the transducer to the CT-7000 S3 and running a transducer self-test using the instructions below: a. Start from the “START-UP” menu: 1. TIME BREAKER 09/19/13 2.
Page 117
REV 1 CT-7000 S3 USER’S MANUAL d. The screen will be updated with the new transducer position information: TRANSDUCER TEST: T1= 5.00 in / 360.0° T2= 0.00 in / 000.0° T3= 0.00 in / 000.0° R1= 0.00 IN R2= 0.00 IN R3= 0.00 IN...
CT-7000 S3 USER’S MANUAL REV 1 Testing the Cable Hookups You can check the CT-7000 S3 contact cable connections to the circuit breaker using the steps below: a. Start from the “START-UP” menu: 1. TIME BREAKER 09/19/13 2. GET RESULTS 10:10:10 3.
Page 119
REV 1 CT-7000 S3 USER’S MANUAL [START] [STOP] Press the key to retest the cable connections, or press the key to end the test and return to the “START-UP” menu.
REV 1 Printing Raw Test Record Data You can print the raw hexadecimal timing data collected by the CT-7000 S3 using the “PRINT DATA” command. This feature is used mainly for factory diagnostic testing. The CT-7000 S3 collects 20,000 data points in the one-second timing window. As a result, each data sample is 0.05 ms apart.
REV 1 CT-7000 S3 USER’S MANUAL d. The following screen will be displayed: 1. EVERY POINT 2. EVERY 5 POINTS Press the key to print every data point. The data points will be printed and you will be returned to the “START-UP” menu.
A dual operation OPEN-CLOSE or CLOSE-OPEN will result in a 0ms reading. • The CT-7000 S3 contact channels may be damaged or the cables may be bad. Try the “Check Hook-UP” test (see section 4.3). No contact time (0 ms) in CLOSE- •...
Page 123
REV 1 CT-7000 S3 USER’S MANUAL When performing an OPEN-CLOSE • Delay between OPEN to CLOSE is probably too short. Increase the test, the breaker opens but will not delay time. close. No data or erratic reading on travel • Bad transducer. Perform a transducer diagnostic test using the transducer.
CT-7000 S3 USER’S MANUAL REV 1 APPENDIX E – CT-7000 S3 Transducer Illustrations Linear Transducer on an OCB Rotary Transducer on Alsthom HE 1014 Circuit Breaker...
Page 133
REV 1 CT-7000 S3 USER’S MANUAL Rotary Transducer on ABB HMB Mechanism Rotary Transducer on Federal Pacific Circuit Breaker...
REV 1 CT-7000 S3 USER’S MANUAL APPENDIX F – CT-7000 S3 Accessories ON-LINE DUAL GROUND TIMING PROBES PROBES (3 probes) part no 9021-OT part no 9021-DG LINEAR ROTARY TRANSDUCERS TRANSDUCER WITH 10" part no 9087-UC MAGNETIC 25" part no 9088-UC...
Page 136
CT-7000 S3 USER’S MANUAL REV 1 DOBLE TRANSDUCER ADAPTER Part no 9084-UC The Doble Transducer Adapter can be used to interface any Doble travel transducer to a Vanguard circuit breaker analyzer. Connected to Doble rotary transducer...
Need help?
Do you have a question about the CT-7000 S3 and is the answer not in the manual?
Questions and answers