5. Ensure the CPR dial is rotated to a closed position.
6. Attach pump power lead to mains power outlet, running power lead along the cable tidy.
7. Switch on.
8. The display will indicate the mattress is inflating, which will complete after approximately 20 mins
9. When the system is ready to use there is an audible signal (beep) and the panel will display 'System is ready to use'
Quick Glance User Guide
The Artemis mattress is a 2 cell alternating Dynamic Replacement
A
with a patient weight limit of 39st (248Kg). It is suitable for patients up
to and including category 4 pressure ulcer with frequent monitoring
and repositioning.
Contraindications: Appropriate advice should be taken for patients with
unstable spinal injuries. Do not use the bed if the patient and accessories
exceed patient weight limit (MUW).
1.
Place mattress directly onto the mattress platform
ensuring umbilical cord is at foot end and 'feet'
symbol is facing up.
2.
Attach straps to moving parts of bed frame
making sure they are not twisted.
3.
Position control unit on foot board of bed using
the integral brackets or on a firm surface.
4.
Attach umbilical cord to control unit by aligning
with the ports on control box and pushing into place.
5.
Ensure the CPR dial is rotated to a closed position.
B
Leaflet Code: LL279
Need Assistance?
C
Call 01422 233 136
(8:30am - 5pm)
or 0800 037 0234
(out of office hours)
FOR COMPLETE FUNCTIONALITY, PLEASE REFER TO THE FULL USER GUIDE
OPERATION
Therapy:
Once the patient is on the mattress, the
mattress will default to alternating therapy.
Select the therapy required by pressing
the therapy button.
Therapy Options:
• Alternating
• Constant Low Pressure
• Pulsation
• Max Inflate – Automatically returns to
6.
Attach pump power lead to mains power outlet,
alternating after 20 mins.
running power lead along the cable tidy.
Once you have chosen the Therapy by using the
7.
Switch on.
up/down arrows you can press the return button
8.
The display will indicate the mattress is inflating,
to confirm your selection and return to the home
which will complete after approximately 20 mins
screen.
9.
When the system is ready to use there is an
If using the CLP and Pulsation therapies to return
audible signal (beep) and the panel will display
to alternating you will need to select Alternating
'System is ready to use'
from the Therapy menu.
The therapy chosen will be displayed on the
home screen.
Comfort Control:
The Feature function enables the carer to make
slight changes to the firmness of the mattress.
The pump defaults to medium and allows the carer
to increase or decrease the pressures in the
mattresses by selecting firm or soft, dependent on
the patients' requirements.
Control Unit Lock / Unlock Function:
The pump will automatically lock if inactive after 2
minutes. Pressing the lock button will unlock the
pump.
Pump Control Unit:
A
B
Operation
Mute Function:
A.
Therapy
Once the patient is on the mattress, the mattress will
Silences the audible alarm. The audible alarm
D
default to alternating therapy.
will resume after 20 minutes if cause of failure is
Select the therapy required by pressing the therapy
unresolved.
button.
The audible alarm also sounds when the power
Therapy Options
is switched off – press alarm mute to silence.
• Alternating
• Constant Low Pressure
E
Home Button: Returns you to the home screen.
• Pulsation
• Max Inflate – Automatically returns to alternating
after 20 mins.
Low Patient Weight: Significantly reduces
Once you have chosen the Therapy by using
F
pressure for low weight ie. patients <40kg.
the up/down arrows you can press the return
button to confirm your selection and return to the
Press for 5 seconds to activate
home screen.
If using the CLP and Pulsation therapies to return to
LED Screen Display:
alternating you will need to select Alternating from
the Therapy menu.
The screen displays information to the carer:
The therapy chosen will be displayed on the
• Weight indication of the patient – should be used as a
home screen.
guideline only.
• Chosen Therapy: Fowler position – if back rest has been raised
B.
Comfort Control
to 30º or SEAT CUSHION if the seat cushion is connected.
The Feature function enables the carer to make
slight changes to the firmness of the mattress.
• Internal cell pressures
• Cycle time
The pump defaults to medium and allows the
carer to increase or decrease the pressures in the
mattresses by selecting firm or soft, dependent on
Egress Alert:
the patients' requirements.
To activate the Egress Alert press 'Comfort Control' followed
C.
Control Unit Lock / Unlock Function
by 'Lock' and 'Down arrow' simultaneously. Choose option 5
The pump will automatically lock if inactive after
and select on/off accordingly.
2 minutes. Pressing the lock button will unlock
the pump.
INSTRUC/ARTEMIS/QUICK - Rev 7 - 09/08/2016
D
E
F
C
D.
Mute Function
Silences the audible alarm. The audible alarm
will resume after 20 minutes if cause of failure is
unresolved.
The audible alarm also sounds when the power is
switched off – press alarm mute to silence.
E.
Home Button
Returns you to the home screen.
F.
Low Patient Weight
Significantly reduces pressure for low weight ie.
patients <40kg. Press for 5 seconds to activate.
G.
Mute alarm button
Mutes audible alarm for 20 minutes.
LED Screen Display
The screen displays information to the carer:
• Weight indication of the patient – should be used
as a guideline only.
• Chosen Therapy: Fowler position – if back rest
has been raised to 30º or SEAT CUSHION if the
seat cushion is connected.
• Internal cell pressures
• Cycle time
Egress Alert
To activate the Egress Alert press 'Comfort Control'
followed by 'Lock' and 'Down arrow' simultaneously.
Choose option 5 and select on/off accordingly.