C-2
Calibration Programs

Introduction

This appendix includes a calibration program written in BASIC to help you calibrate the
Model 2425, as well as an example program that demonstrates how to request calibration
constants. Refer to Section 2 for more details on calibration procedures, equipment, and
connections. Appendix B covers calibration commands in detail.

Computer hardware requirements

The following computer hardware is required to run the calibration programs:

Software requirements

In order to use the calibration programs, you will need the following computer software:

Calibration equipment

The following calibration equipment is required:
See Section 2 for detailed equipment specifications.
IBM PC compatible computer.
Keithley KPC-488.2, KPS-488.2, or KPC-488.2AT, or CEC PC-488 IEEE-488
interface for the computer.
Two shielded IEEE-488 connecting cables (Keithley Model 7007).
Microsoft QBasic (supplied with MS-DOS 5.0 or later).
MS-DOS version 5.0 or later.
HP-style Universal Language Driver, CECHP.EXE (supplied with Keithley and
CEC interface cards listed above).
Hewlett-Packard HP3458A Digital Multimeter
1Ω, ±0.1%, 100W resistor
Model 2425 Service Manual