3.1 BASIC
Measurement setting
Please go to measurement page to do measurement settings. There is a "Tools" icon in the
lower left cornera. Note that this icon is the Measurement Tools Icon as opposed to the
System Tools icon on the Home Screen.
Previous
Back to
page
measurement page
1
2
This option is turned on can
When saving data (e.g.
be file naming when you
from the BASIC screen), you
saving measured data.
can choose to save to an
Excel file, or Excel & JPG
files (screen shots).
1.
Continuous capture function allows users to take consecutive measurements and review
results quickly without bothering with buttons - however, it does NOT allow you to save data to
the SD Card. (see the LOG function for continuous captures with data-save to an SD Card).
2.
Live view mode only for BASIC, SPECTRUM, CIE1931, CIE1976, CRI measurements.
Next
page
3
4
Integration time can be
This option is turned on in
automatically set by the MK350S
low light conditions (LUX ≦
Premium (Auto Mode) or
50). It allows you to set the
Manually (Manual Mode). If
Integration time (exposure
Manual Mode is chosen,
time) higher.
you must set "Integration
time"
- 19 -
I-Time
Setup
Auto
a
Off
b
60
μs
~
500
ms
c
A
A L
-Automatic+Long Time Exposure
-Automatic
M
M L
-Manual
-Manual+Long Time Exposure
5
6
The length of time, in µs,
You can set the MK350S
that the MK350S Premium
Premium to take snapshots
captures light data in one
continuously
(Continuous
measurement (like exposure
capture). As soon as you press
time).
the "Measure" button, the unit
will begin to capture at 3 second
intervals. Press "Measure"
again to stop capturing.
a
b
Auto
Manual Manual
c
The system will prompt you to do Dark
On
Off
On
calibration after you turn on/off "Long
exposure time". This procedure will take you
around 1min and 15 secs. We strongly
60
60
60
μs
μs
μs
~
~
~
recommend that you to execute Dark
5000
500
5000
ms
ms
ms
Calibration every time you on/off "Long time
exposure" setup.
7
8
This turns on and off the
This item gives you the option
audible beep that occurs
of reviewing the viewport
when you take a measurement
before you take a measurement
- it takes up to 3 seconds
with the "Measure" button.
take a measurement – the
unit beeps when it is done.
- 20 -
9
You are always prompted to
do a Dark calibration when
you
turn
on
the
unit.
However, you can perform a
dark
calibration
anytime
with this option.