Do not connect the USB cable in ways other than those mentioned in this manual. ◗ ◗ counting. With this, the Corning Cell Counter can accurately and precisely measure cell If water or other liquid enters the device, discontinue use immediately. Continued use ◗...
2.0 Getting Started Instrument Overview The hardware of the Corning® Cell Counter is shown below (Figure 1): Product contents include the following items (Figure 2): 1. Light arc – Contains the light source for imaging 1. Manual and Quick Starter Guide 2.
App on the Windows computer. Figure 6. Sign up registration form 3. Be sure to remove any wrapping before first time use of the Corning Cell Counter. Also clean the device with a damp, lint-free wipe to clean the outer case with 70% ethanol or isopropanol.
Is your personal account linked to more than one company? In that case, you will see With the Corning® Cell Counter you can perform a single count or multiple counts the comment box as shown in Figure 8 after signing in with your email address and within a project.
For proper analysis, you should be seeing a contrast: living cells having bright centers The optical window of the Corning Cell Counter must be clean. You can check this by and dark edges (Figure 18). To achieve this, using the focus knob, set the focus just monitoring the live screen.
Performing a Count When the focus is set, you are ready to perform a count. To obtain your cell count, follow One of the features of the Corning® Cell Counter is that the experiment results can be the protocol below: accessed through the CytoSMART™...
This link will lead you to the CytoSMART™ Connect login page (Figure 21). Log in using Experiment Overview the credentials you created earlier. If you have multiple customer environments, choose After selecting a project (C), you will get access to an overview of all experiments which environment you would like to access (Figure 22).
70% ethanol or isopropanol. If you see dust particles on the optical window, remove them by cleaning with a lint-free wipe. Important: Do not autoclave the Corning Cell Counter, as this can result in damage to the electronics. The Corning Cell Counter does not need regular maintenance. Do not perform any repairs or service on the device yourself.
Corning® CytoSMART™ Cloud App, you may Cell Counter (Figure 27). To resolve this, pellet the solution by see bubbles (Figure 30). They are centrifuging and resuspend the cells in a smaller volume.
This warranty gives you specific legal rights. You may have other rights which vary from state to state. No individual may accept for, or on behalf of Corning, any other obligation of liability, or extend the period of this warranty.
According to Directive 2012/19/EU of the European Parliament and of the Council of 4 July 2012 on waste electrical and electronic equipment (WEEE), the Corning Cell Counter is marked with the crossed-out wheeled bin and must not be disposed of with domestic waste.
Page 15
L A T I N A M E R I C A t 886 2-2716-0338 grupoLA@corning.com f 886 2-2516-7500 Brasil t 55 (11) 3089-7400 Mexico t (52-81) 8158-8400 For a listing of trademarks, visit www.corning.com/clstrademarks. All other trademarks are the property of their respective owners.
Need help?
Do you have a question about the Cell Counter and is the answer not in the manual?
Questions and answers