W
ARNINGS
Perform Service procedures only in an ESD-protected environment.
This device is not intended for life support.
Do not service the Alice PDx device in a Magnetic Resonance Imaging (MRI) environment or
in close proximity to a high emissions source.
The Alice PDx system and software are not designed to replace the clinical judgement and
analysis of a health care professional.
Do not plug sensor cables into electrical outlets. Cable contact with electrical outlets presents
a serious shock hazard.
Make sure the conductive parts of electrodes and associated connectors, including the neutral
electrode, do not contact other conductive parts, including earth.
Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of
an open flame.
Do not service the device in the presence of a flammable anaesthetic mixture in combination
with oxygen or air, or in the presence of nitrous oxide.
When attaching the sensors and cables, be careful to route the cables in a manner that will
reduce the possibility of damage to either the sensors or the device.
Periodically inspect the sensor cables for damage or signs of wear. Replace if damaged.
Do not immerse the Alice PDx in any fluids.
Repairs and adjustments must be performed by Respironics-authorized service personnel
only. Unauthorized service could cause injury, invalidate the warranty, or result in costly
damage.
Use caution when removing damaged batteries and avoid exposing skin to any battery
leakage.
Recycle or dispose of batteries in accordance with local regulations. Do not incinerate.
Use only accessories that have been approved by Respironics.
Connecting the Alice PDx to a device that is not approved by Respironics with the therapy
and/or SleepLink
devices that are IEC 60601-1 approved may be attached to the Alice PDx.
Ensure that any computer connected to the Alice PDx complies with the safety standard IEC
60950. Only connect the Alice PDx to an IEC 60950 compliant computer when configuring the
device or when viewing a sleep study in real time, carefully following the respective
instructions. Do not connect the Alice PDx to any other USB compatible device.
Do not use grease, oils, polishes or other petroleum-based products when servicing or
handling the device.
Precautionary procedures include methods to prevent buildup of electrostatic discharge (e.g.,
air conditioning, humidification, conductive floor coverings, and non-synthetic clothing),
discharging one's body to the frame of the equipment or system or to earth or a large metal
object, and bonding oneself by means of a wrist strap to the equipment or system, or to earth.
P
4
AGE

WARNINGS

®
communication cables, may result in a shock hazard to patient. Only
TM
®
Alice
PDx
Service & Technical Reference Manual
1054417, V
. 01
ER