B
EPLACING THE
ATTERY
To replace the Battery Enclosure:
1. Using a Torx T8 screwdriver, remove and retain the four screws that secure the Battery Enclosure
to the Clamshell Assembly.
2. Lift the Battery Enclosure away from the Battery PCA/Battery Enclosure Cover Assembly. Refer to
the following illustration.
Battery Enclosure
The Flex Cable is
Delicate. Handle
with caution.
To install the Battery Enclosure:
1. Be sure the Battery PCA is properly seated in the Battery Enclosure Cover.
2. Align the Battery Enclosure with the Battery PCA/Battery Enclosure Cover Assembly and properly
seat the Battery Closure in place.
3. Secure the Battery Enclosure with the four screws previously removed.
1054417, V
. 01
ER
E
NCLOSURE
2.5 in.-lbs. (x4)
Battery PCA
F
C: B
IGURE
ATTERY
TM
®