Maico MI 34 Operating Instructions Manual page 49

Hide thumbs Also See for MI 34:
Table of Contents

Advertisement

Operating Instructions MI 34 / MI 34 H
To clean the small acoustic and air pressure channels of the probe tip
unscrew the small ribbed plastic nut that holds the probe tip:
Figure 39:
After unscrewing the small ribbed plastic nut it is possible to detach the
small probe tip with the small acoustic and air pressure channels from the
transducer house:
Figure 40:
Ribbed Plastic Nut
Transducer House
Transparent Sealing
Probe Tip with the small acoustic and air pressure channels
The cleaning of the acoustic and air pressure channels of the probe tip must
be performed by means of the cleaning wire which can be found in the Ear
tips Assortment provided with the MI 34.
When cleaning the acoustic and air pressure channels of the probe tip the
cleaning wire must be inserted from the back of the probe tip according to
Figure 41: