Iron, Total,
continued
®
Using AccuVac
1.
Fill a 10-mL cell to the
10 mL line with sample. Cap.
Collect at least 40 mL of
sample in a 50-mL beaker.
Ampuls
2.
Fill an Iron TPTZ
Reagent AccuVac Ampul
with sample.
Note:
ampule immersed until the
ampule fills completely.
Keep the tip of the
1—22
3.
Invert the ampule several
times to mix. Wipe off any
liquid or fingerprints.