WYLER ML NT 11 Operator's Manual

WYLER ML NT 11 Operator's Manual

Classic series; nt series

Advertisement

WYLER AG
Im Hölderli
CH-8405 WINTERTHUR
Switzerland
LEVELTRONIC „classic" and "NT"
MINILEVEL "classic" and "NT"
File: „nt_eng.doc"
O
PERATOR
LEVELMETER 25 / C25
LEVELMETER 2000
Tel.
0041 (0) 52 233 66 66
Fax.
0041 (0) 52 233 20 53
Homepage: http://www.wylerag.com
E-Mail:
wyler@wylerag.com
'
M
S
ANUAL
25.7.1999

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ML NT 11 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for WYLER ML NT 11

  • Page 1 WYLER AG Tel. 0041 (0) 52 233 66 66 Im Hölderli Fax. 0041 (0) 52 233 20 53 CH-8405 WINTERTHUR Switzerland Homepage: http://www.wylerag.com E-Mail: wyler@wylerag.com PERATOR ANUAL LEVELTRONIC „classic“ and “NT” MINILEVEL “classic” and “NT” LEVELMETER 25 / C25 LEVELMETER 2000 File: „nt_eng.doc“...
  • Page 2: Table Of Contents

    Differential respectively reference measurement using an engineer set 1.5.4 Angular measurements 1.5.5 Lines and flatness measurement (manual procedure) 1.6. WYLER software LEVELSOFT 2. DETAILED INFORMATION to the various instruments 2.1. MINILEVEL / LEVELTRONIC „classic“ 2.1.1 MINILEVEL “classic” A10 / LEVELMETER C25/DC (red cover) 2.1.2...
  • Page 3: Index

    Change of Measuring Unit, with Minilevel “NT” 2.2.3.7 Classic, Instruments Collecting the data from a LM 2000 to a PC when using WYLER sw LEVELSOFT 2.2.3.8 Collecting the data from a LM 2000 to a PC when using WYLER sw LEVELSOFT 2.2.6.8...
  • Page 4 LEVELMETER C25/DC with Minilevel “ / Leveltronic NT” 2.2.2 LEVELMETER 2000 with Minilevel “NT” 2.2.3 LEVELMETER 2000, overview LEVELSOFT PRO, WYLER software LEVELTRONIC “classic” / A40 LEVELTRONIC “classic” A40 / LEVELMETER 25 2.1.4 LEVELTRONIC “NT” / 41 LEVELTRONIC “NT” 41 with LEVELMETER C25/DC (red cover) or LM2000 2.2.4...
  • Page 5 Sensitivity of the instrument 1.5.5 Small angles 1.5.4 SPARE PARTS / ACCESSORIES / Square angles 1.5.4 Start the LEVELMETER 2000 2.2.3.1 Start the LEVELMETER 2000, with Leveltronic “NT” 2.2.6.1 STORING Switch off the Levelmeter 2000 2.2.6.9 Switch the Levelmeter 2000 off 2.2.3.9 TECHNICAL DATA UNION JACK...
  • Page 6: Basics

    Instrument's overview Series „classic“ Series „NT“ Minilevel A10 „classic“ with integrated display ML NT 11 with integrated display Leveltronic A40 “classic” Leveltronic NT 41 with LEVELMETER 2000 Engineer set consisting of 2 Leveltronic A40 „classic“ 2 Leveltronic NT 41 with LEVELMETER 2000...
  • Page 7: Measuring Instruments / External Display Instruments

    MINILEVEL „classic“ for use with LEVELMETER C25/DC (red cover) / Art.-no. WYLER: 025-005 MINILEVEL „NT“ with LEVELMETER 2000 / Art.-no. WYLER: 065-004-001 or MINILEVEL NT 11 LEVELMETER C25/DC (red cover) / Art.-no. WYLER: 025-005 The LEVELTRONIC instruments are not equipped with an internal display and require therefore the connection to an external display unit.
  • Page 8: Measuring Procedure / General Use

    Measuring procedure / general use 1.3.1 Basic set-up of the inclination measuring instruments / plus - minus rule The instrument should always be placed in a way that the cable output is on the right hand side. Inclined to the right: positive Declined to the right: negative (Practical remark: positive values indicate that some material may be...
  • Page 9: Absolute Measurement / Relative Measurement / Differential Measurement

    1.3.2 Absolute measurement / Relative measurement / Differential measurement Absolute measurement (Absolute ZERO) Condition for the absolute measurement is the performing of a reversal measurement for determining the absolute zero (to the centre of the earth) After this procedure the instrument will display the effective deviation from the centre of gravity.
  • Page 10: General Remarks About "Angles

    1.3.3 General remarks about "ANGLES" Every angle may be defined in different ways. The most popular way is a definition in Degrees / Minutes / Seconds. This is shown in the graph. Such a definition is especially useful for larger angles. With an angular measuring instrument not only an angle may be measured but also the height of a certain point over a defined base length can be...
  • Page 11: Height Related To The Step Length

    1.3.4 Height related to the step length × Height µm Steplength Height Hx in [µm] related to Height Hx [µm] = Length step length × Height mikroINCH Steplength Inch Height Hx in [micro Inch] related Height Hx [micro Inch] = Length Inch to step length...
  • Page 12: Zero Setting By Reversal Measurement (Absolute Zero)

    It is not always required to set the instrument to zero by reversal measurement (absolute ZERO). If e.g. the straightness of a line is measured only the relative most of the time the relative straightness is only required. The same applies when lines, parallels, flatness is measured with the WYLER LEVELSOFT. These are relative measurements.
  • Page 13: Differential Respectively Reference Measurement Using An Engineer Set

    Picture: Two LEVELTRONIC „NT“ and a LEVELMETER 2000 connected to laptop equipped with the WYLER software LEVELSOFT Remarks concerning NT instruments: It makes sense to combine a LEVELTRONIC NT and a MINILEVEL NT to an engineer set. With such a combination a great variety of other measuring tasks requiring a stand-alone instrument may be fulfilled.
  • Page 14 Examples for differential / reference measurement: = Reference instrument, connected to port „-B“ = Measuring instrument, connected to port „A“ Fig.7 Rectangularity resp. parallelism of machine spindle in resp. to work table. For precise square measurements an angular model fitted with magnets in the vertical face should be used. (Elimination of deformation due to manual pressure).
  • Page 15: Angular Measurements

    After the reversal measurement precision measurement with measuring uncertainty of 3 to 5 µm/m is possible. The determination of the instrument's angular error is done as follows: In the WYLER LEVELSOFT 99 a new set-up allows the easy measurement of squareness of rectangular objects.
  • Page 16: Lines And Flatness Measurement (Manual Procedure)

    Ask your local WYLER representative for further information of the software LEVELSOFT 1.5.5 Lines and flatness measurement (manual procedure) A principal procedure of a measurement of a line with WYLER instruments is described below: Example: Sensitivity of the instrument 1 µm/m...
  • Page 17: Wyler Software Levelsoft

    WYLER software LEVELSOFT PRO The WYLER software LEVELSOFT PRO well proven package based on the ISO 1101 for measuring lines and surface flatness constantly upgraded to the demands of the user. The following measurements can be done with the WYLER LEVELSOFT PRO in combination with the correct measuring instruments: •...
  • Page 18 Rectangularity according to ISO1101 In order to connect the measuring instruments with a PC or Laptop an adapter is required • LEVELADAPTER SET 2000 (No. 24, for all type of instruments) or • the special connecting set (Levelsoft no. 24A) to be used with instruments of the NT family together with a LEVELMETER 2000 and the respective cable set.
  • Page 19: Detailed Information To The Various Instruments

    Two of the instruments can also be used for differential measurements in combination with a LEVELMETER C25/DC (red cover) / Art. no. WYLER: 025-005 MINILEVEL „classic“ A10 IMPORTANT: The batteries must be removed in the MINILEVEL "classic"...
  • Page 20: Levelmeter C25 / Dc

    2.1.2 LEVELMETER C25 / DC Switch (Backside) Switch for LEVELMETER. The low battery indication Top view "On" / "Off" (XX:XX colon) is appearing automatically during LEVELMETER C25 / DC operation if battery power is insufficient. Potentiometer: Enables adjustment of the zero point. The regulating "ZERO"...
  • Page 21 Possible configurations with MINILEVEL „classic“ MINILEVEL „classic“ has an integrated display and can be used as a stand alone instrument MINILEVEL „classic“ with LEVELMETER C25 (red cover) Two MINILEVEL „classic“ with LEVELMETER C25 (red cover) for differential measurement Function check of the system MINILEVEL „classic“ combined with LEVELMETER C25 LEVELMETER C25 / DC power ON without cables connected Low battery indication (XX:XX colon) must not appear Connect MINILEVEL „classic“...
  • Page 22: Leveltronic "Classic" A40 / Levelmeter 25 (Blue Cover)

    2.1.3 LEVELTRONIC “classic” A40 / LEVELMETER 25 (blue cover) The LEVELTRONIC „classic“ can only be used in connection with a LEVELMETER 25 (blue cover) , Only one measuring range is available. LEVELTRONIC „classic“ A40 2.1.4 LEVELMETER 25 Switch (Backside) Switch for LEVELMETER. The low battery indication Top view "On"...
  • Page 23 Possible configurations with LEVELTRONIC „classic“ LEVELTRONIC „classic“ combined with LEVELMETER 25 (blue cover) Two LEVELTRONIC „classic“ combined with LEVELMETER 25 (blue cover) for differential measurement Function check of the system LEVELTRONIC „classic“ combined with LEVELMETER 25 LEVELMETER 25 power ON without cables connected Low battery indication (XX:XX colon) must not appear Connect LEVELTRONIC „classic“...
  • Page 24: Minilevel / Leveltronic „Nt

    Two of the instruments can also be used for differential measurements in combination with a LEVELMETER C25/DC (red cover) / Art. no. WYLER: 025-005 or to a LEVELMETER 2000 / Art. no. 065-004-001 MINILEVEL „NT“ 11 MINILEVEL NT 11 The MINILEVEL NT has three pushbuttons.
  • Page 25: Levelmeter

    2.2.2 LEVELMETER C25/DC Switch (Backside) Switch for LEVELMETER. The low battery Top view "On" / "Off" indication (XX:XX colon) is appearing LEVELMETER C25 / DC automatically during operation if battery power is insufficient. Potentiometer: Enables adjustment of the zero point. The "ZERO"...
  • Page 26: Levelmeter 2000

    2.2.3 LEVELMETER 2000 2.2.3.1 Start the LEVELMETER 2000 press, ON/MODE key. The measuring mode last used is now being restarted, then a display appears. 2.2.3.2 Measurement absolute (1 Instrument only ) press repeatedly, until the cursor is positioned under SENSOR confirm selection.
  • Page 27 Displayed mode changes to ABSOLUTE 2.2.3.5 Measuring with relative Zero (1 instrument only) Select channel „A or B“ see 2.2.3.2 Set instrument to relative ZERO press repeatedly until the cursor flashes under REL ZERO confirm. Direction indicator is flashing. The last stored Offset is displayed. Put the instrument in the designated position.
  • Page 28 2.2.3.8 Collecting the measured data from a LEVELMETER 2000 to a PC when using the WYLER software LEVELSOFT When using the Software Levelsoft in combination with a LEVELMETER 2000 the key <ENTER> has the identical function as the push button of the connecting cable (Dongle cable) or the push button of the infrared sender unit for data collection.
  • Page 29 Possible configurations with MINILEVEL „NT“ 11 MINILEVEL NT has an integrated display and can be used as a stand alone instrument Analogue data transmission Digital data transmission (RS485) MINILEVEL NT with MINILEVEL NT with LEVELMETER C25 LEVELMETER 2000 (red cover) Two MINILEVEL NT Two MINILEVEL NT with LEVELMETER C25 (red cover)
  • Page 30 Zero setting of the measuring system MINILEVEL „NT“ 11 with LEVELMETER C25/DC Prior to adjusting the zero point, the instrument should be allowed time to acquire the ambient, respectively the work-piece temperature. Also the measuring system should have been switched on for about 1 min. (for sensitivity of 1 µm/m about 3 minutes) The absolute zero setting can be done in two ways.
  • Page 31: Leveltronic "Nt" 41 With Levelmeter C25/Dc (Red Cover) Or

    LEVELTRONIC “NT” 41 with LEVELMETER C25/DC (red cover) or LEVELMETER 2000 The LEVELTRONIC „NT“ 41 can only be used in combination with a LEVELMETER C25/DC (red cover) Art. no. WYLER: 025-005 or a LEVELMETER 2000. Art.-no. 065-004-001. It is also possible to connect two instruments to the mentioned Levelmeter for differential measurements.
  • Page 32 2.2.5 LEVELMETER C25/DC Switch (Backside) Switch for LEVELMETER. The low battery Top view indication (XX:XX colon) is appearing LEVELMETER C25 / DC "On" / "Off" automatically during operation if battery power is insufficient. Potentiometer: Enables adjustment of the zero point. The "ZERO"...
  • Page 33: Levelmeter 2000

    2.2.6 LEVELMETER 2000 2.2.6.1 Start the LEVELMETER 2000 press, ON/MODE key. The measuring mode last used is now being restarted, then a display appears. 2.2.6.2 Measurement absolute (1 Instrument only ) press repeatedly, until the cursor is positioned under SENSOR confirm selection.
  • Page 34 2.2.6.5 Measuring with relative Zero (1 instrument only) Select channel „A or B“ see 2.2.3.2 Set instrument to relative ZERO press repeatedly until the cursor is flashing under ZERO confirm (Direction indicator flashing) and the previously stored ZERO offset is displayed.
  • Page 35 2.2.6.8 Collecting the measured data from a LEVELMETER 2000 to a PC when using the WYLER software LEVELSOFT When using the Software Levelsoft in combination with a LEVELMETER 2000 the key <ENTER> has the identical function as the push button of the connecting cable (Dongle cable) or the push button of the infrared sender unit for data collection.
  • Page 36 Possible configurations with LEVELTRONIC „NT“ 41 Analogue data transmission Digital data transmission (RS485) LEVELTRONIC „NT“ 41 with LEVELMETER C25 LEVELTRONIC „NT“ 41 with (red cover) LEVELMETER 2000 Two LEVELTRONIC „NT“ 41 with LEVELMETER C25 Two LEVELTRONIC „NT“ 41 with LEVELMETER 2000 (red cover) for differential measurement for differential measurement Function check of the system LEVELTRONIC NT 11 combined with LEVELMETER C25/DC...
  • Page 37 ZERO Setting (absolute) of the LEVELTRONIC „NT“ 41 with LEVELMETER 2000 by reversal measurement Prior to adjusting the zero point, the instrument should be allowed time to acquire the ambient, respectively the work-piece temperature. Also the measuring system should have been switched on for about 1 min. (for sensitivity of 1 µm/m about 3 minutes) Condition: Only one channel "A"...
  • Page 38: Technical Data

    TECHNICAL DATA LEVELTRONIC “classic” / A40 Sensitivity, 1 Digit = 1 µm/m 5 µm/m 10 µm/m 0.2 Arcsec 1 Arcsec 2 Arcsec Measuring range +/- 1999 Digits 2 mm/m 10 mm/m 20 mm/m 400 Arctic 2000 Arcsec 4000 Arcsec Limits of error <0.5 Full scale value (DIN 2276) max.
  • Page 39: Maintenance

    Below you will find a "trouble-shooting" list. If this does not help you to remove your difficulty the instrument must be sent to your nearest WYLER service engineer. When repairing the electronic unit, special care must be taken as the integrated circuits are sensitive to static charges.
  • Page 40 Difficulty: Zero point unattainable - but instrument operates normally (Sensitivity) Possible cause: The instrument may have been subjected to a severe shock the zero point is dislocated in the sensor head. Remedy: ...LEVELTRONIC „classic“ A40 with LEVELMETER 25 1. Check the measuring base for damage. If damage is ascertained the instrument must be sent for re- conditioning.
  • Page 41: Leveltronic "Nt

    All the MEASURING INSTRUMENTS are available with a great variety of measuring bases. To the LEVELMETER the following parts are available: Mains adapter WYLER No. 29 for connection to 220V/50HZ also possible to be used for charging rechargeable batteries. Other voltage adapters on request.
  • Page 42: Repair Of Measuring Instruments

    The invoicing will be through the WYLER partner in your country. Please do not hesitate to make use of this service in order to have your WYLER instrument back at your disposal as soon as possible. In case of any questions please contact WYLER SWITZERLAND or your local distributor, we will gladly help you to use the ERS successfully.

Table of Contents