Preparation Before Usage - Haier HYC-260 Operation Manual

Pharmaceutical refrigerator
Table of Contents

Advertisement

Preparation before usage

1. Remove all packaging materials.
For safety reason, the equipment of
HYC-890/940/940F is situated on a wooden
pallet and secured by metal brackets. Please
remove the screws and place the metal
brackets under the unit.(Pic.1).
Forklift or specialty lifting equipment should be
used to remove the unit off the pallet. Forklift
should reach the bottom of the wood pallet for
lifting.
The unit should not be tilted at an angle of
more than 45 degree.
2. Inspect standard accessories shipped with the unit
Inspect the accessories against packing list. Should
there be any discrepancy, contact the after-sale
department.
3. Installation
Leave a minimum space of 10 cm around the
refrigerator for ventilation and heat dissipation.
(Pic.2).
4. Adjust the leveling leg
Rotate the leveling legs with a wrench in clockwise
to extend them out and secure them onto the ground.
This is to make sure the refrigerator does not move
during operation.(Pic.3).
8
Pic.1
> 30cm from the top
> 10cm from
> 10cm from
the right
the left
Pic.2
Downward
upward
Leveling legs
Pic.3
Circuit diagram(HYC-890)
N
37

Hide quick links: