V13026 Appearance , color and layout of the door may vary. Translation of the original instruction. Haier pharmaceutical refrigerators are applicable for the storage of pharmaceutical products, medicines, and other laboratory products that need to be stored at a 0270501000...
0.129 HYC-118A HYC-118A 220~240 50/60 0.122 Manufactures Name : Haier Medical and Laboratory Products Co.,Ltd. into the atmosphere. Manufactures Address : Haier Industrial Park, Economic Technology GWP=global warming potential Development Zone. Qingdao 266510.P.R.China Refrigerant type Refer to in this declaration confirms with the following directive(s)/standards:...
Page 3
Name Pharmaceutical Refrigerator HYC-118A Model HYC-118 Exterior Dimensions The product is applicable to pharmacies, pharmaceutical factories, quarantine stations, health 597×635×835 (W×D×H) 597×635×835 centers and hospitals, used to store biological products and those need to be stored at a (mm) temperature between 2~8℃. Interior Dimensions (W×D×H) 515×415×630...
Safety Precautions Dear users: Thanks for your choice of Haier pharmaceutical refrigerator; please make sure you have carefully Circuit diagram read and observed the contents with following signs in the manual, for better understanding of this manual and better use of this product, so as to prevent personal injuries and refrigerator damage.
Page 6
Refrigeration Schematic & Wiring Diagram Do not place containers with water or heavy stuffs on the refrigerator. Personal injuries Schematic diagram may be caused by falling articles and electrical leakage or shock may be caused by out flowed water due to deterioration of insulation Do not ground the refrigerator through gas pipes, water supply pipes, telephone lines or lightning conductors as electric shock or other dangers may be caused.
Page 7
Check refrigerator settings when restarting the refrigerator after power failure or the still unresolved after operation, please contact Haier after-sales service.Do not power is shut off. Changes of settings may damage the articles stored.
Battery switch “I” indicates that battery is on ; old appliances with new ones, the retailer is legally obligated to take back your old appliance for “O” indicates that battery is off。 disposals at least free of charge. HYC-118/118A HYC-68/68A...
Product Installation 5.Check alarm temperature: See Fig #1 Installation environment Press both “Set” and “+” buttons to check the most recent cabinet temperature. Release the buttons to return to the cabinet temperature display. Ambient temperature: 16℃ to 32℃, 18℃ to 25℃ are optimal and air conditioning system is Press both “Set”...
Application method Preparation before use 1.Microprocessor temperature controller Adjust the cabinet temperature easily by the micropressor temperature controller in control panel 1. Remove the packaging materials and strings installed on the top of the refrigerator. Don’t do any adjustment in the first 30 mins when you first Remove all packaging materials and strings for transport.
HYC-68A HYC-118A Initial Power-on Observe the rules below for initial start-up and continuous operation: Door lock Control panel Inner light Door lock Control panel Inner light When the refrigerator is empty, connect the power line to the appropriate dedicated socket and make sure the power is rated voltage and frequency.
This unit should be managed by a dedicated person. Inspection should be performed daily to keep appropriate records. When the temperature is out pf specification, action should be taken to move the products to a safe place. HYC-68 HYC-118 This product is pharmaceutical refrigerator with temperature ranging from 2~8℃ inside.