Circuit Diagram

Circuit diagram(HYC-260/360)
N
34
Initial Power-on
Follow the flow chart below to start the refrigerator.
While the refrigerator is empty, connect the power line to the appropriate dedicated socket and
make sure the power supply in the socket matches the required voltage and frequency.
After the power is switched on, turn on the battery charging switch. The audible alarm is usually
triggered. Press the buzzer to stop the alarm. The audible alarm continues to operate until the
temperature of the sensor reaches 5℃±3℃.
The refrigerator has been preset to operate at 2~8℃ in the factory and it is not required to set
temperature
It usually takes several hours for the refrigerator temperature to reach a stable condition. When
the temperature becomes stable, inspect the actual temperature to make sure it matches the set
point.
Turn on the light switch and ensure the light inside the freezer operates properly.
When the refrigerator passes all the functional inspections, load in products in batches.