Haier HYC-260 Operation Manual page 18

Pharmaceutical refrigerator
Table of Contents

Advertisement

HYC-610
Light switch
Glycerin bottle for
temperature sensor
Handle
Door lock
Caster
16
Control panel
Interior light
Shelf
Door ajar alarm switch
USB port(Optional)
Temperature
recorder
Leveling leg
Alarm
Alarm or Safety
Phenomenon
If upper temperature sensor or lower
High
temperature sensor perceives
temperature
temperature of the refrigerator ≥8℃
If upper temperature sensor or lower
Low
temperature sensor perceives
temperature
temperature of the refrigerator < 2℃
Outage of the refrigerator
Outage
Door ajar
Outer door ajar or open
Open circuit or short circuit of
the upper temperature sensor
Open circuit or short circuit of
the lower temperature sensor
Sensor
Open circuit or short circuit of
abnormities
the control sensor
Open circuit or short circuit of
the defrosting sensor
If alarm condition is not corrected within 20 minutes, buzzer and remote control
contact will restart.
IF there is a power outage, a fully charged emergency battery can keep the
alarm function in working condition for 48 hours.
Once the unit starts up and restarts after a long time in storage, charge the
battery for to keep the alarm function for 48 hours.. Even if the battery
discharges, remote alarm contact shall keep in working condition.
Press "Alarm Test " key. For each press, the buzzer goes off thrice with frequency
as 1Hz continuously. Meanwhile, the alarm indicator light flashes thrice and the
remote alarm relay is disconnected after pull-in thrice, which means the alarm
system functions normally.
Alarm Indication
Buzzing Alarm
Alarm light flashes
Impulsive sound alarm
Alarm light flashes
Impulsive sound alarm
Temperature Display area
Impulsive sound alarm
displays inside temperature
within 48h outage
for 60s and stop displaying
for 60s, then repeat
Give impulsive sound
Alarm light flashes
alarm after 10 min delay
after 10 min delay
Alarm light flashes,
Impulsive sound alarm
temperature display
area displays E1
Alarm light flashes,
temperature display
Impulsive sound alarm
area displays E2
Alarm light flashes,
Impulsive sound alarm
temperature display
area displays E3
Alarm light flashes,
temperature display
Impulsive sound alarm
area displays E4
29