Error codes

Display code
Explanation
- 0 -
Door is open
A1
Door alarm "dAd" has been activated
A2
High temperature alarm "HL" is or has been activated
A3
Low temperature alarm "LL" is or has been activated
F1
Error on the main cabinet sensor. The refrigeration system will use an
emergency program to make the cabinet run. Temperature stability will be
affected. Service is required
F2
Error on the evaporator sensor. Service is required
In order to assure the safety of the stored items, the alarms
should be supported by external alarms. This can be done by
utilizing the possibility of the voltage-free contact.
Please find instructions for connecting the "voltage-free contact" in the
"Installation" section.