Toshiba LF654 Instruction Manual page 37

Lf654 series electromagnetic flowmeter detector
Hide thumbs Also See for LF654:
Table of Contents

Advertisement

7.2.3 Flow rate indication is not stable
START
Is power supply voltage
within the specified range?
Are cables securely connected?
Is the flowmeter grounded
with 100Ωor less ground
resistance?
Is the detector pipe filled
with fluid?
Is the fluid carrying bubbles?
Is there high-voltage or large
current cable or equipment
near the flowmeter?
Contact your nearest Toshiba
representative.
Phone: 800.894.0412 - Fax: 888.723.4773 - Web: www.clrwtr.com - Email: info@clrwtr.com
NO
YES
NO
YES
NO
YES
NO
YES
YES
NO
YES
NO
− 36 −
Use a power supply within the
specified range.
Connect each cable securely to the
terminal board.
Ground the flowmeter with a copper
braid or wire(5.5 mm² minimum) to a
good earth ground
Ω
(100
or less ground resistance)
Design piping so that the detector
pipe is filled with fluid all the time.
Design piping so that the fluid does
not carry bubbles.
Install the flowmeter away from
equipment producing electromagnetic
or electro- static interference.
Note 1: If the detector tube is not filled with
operating fluid, the flow is indefinite and
measurement is impossible. Be sure to fill
the detector tube with operating fluid
before starting measurement.