6F8A0926
5.4.2 Cable Connection
Connect and install the terminal-treated cables to the terminal block.
*Connect the cables to the terminal block securely. A loose connection may cause incorrect
measurement. After connecting a cable, try to pull it to check whether it has been connected securely.
(1)Referring to combined converter's manuals of "Connections and Grounding", connect each cable to
the terminal block. Tighten the screws of the terminal block tightly to ensure a secure connection. A
loose connection may cause incorrect measurement. After connecting a cable, try to pull it to see whether
it has been connected securely.
Phillips screwdriver
* The appropriate torque for tightening
the terminal board screw is 1.2 N.m.
Terminal block
Figure 5.5 Connecting a Cable to Terminal Block
(2) Attach the terminal box cover . To keep the seal, tighten securely the cover.