• press Ok to modify the sampling rate (in seconds) of the oximetry parameters. The default rate is every 12 seconds.
Select the required value (4, 8, 12, 16 or 20 sec.) with " ∆ ∆ ∆ ∆ " or " ∇ ∇ ∇ ∇ " . Press Ok to confirm.
6.1.5
Date Format
DATE FORMAT
Esc
•
press Ok to modify the date format. Select the required format (mm/dd/yy or dd/mm/yy) with the " " key. Press Ok to
confirm.
6.1.6
Change Date/Time
CHANGE DATE/TIME
Esc
•
press Ok to modify the DATE and TIME. Use " ∆ ∆ ∆ ∆ " and " ∇ ∇ ∇ ∇ " to modify the number with the cursor
Use the " " key to return to the value previously inserted.
The actual date is shown :
DATE 24
dd / mm / yy
The arrow
positioned to the right of the field, indicates the numerical value to be modified.
WARNING
On the second line under the date format is indicated (mm/dd/yy or dd/mm/yy). If you wish to change the date
format please refer refer to the section "DATE FORMAT"
With the two keys " ∆ ∆ ∆ ∆ " and " ∇ ∇ ∇ ∇ " you increase or decrease the value in the field selected.
In the case that you make an error press " " to move the cursor back to the previous field and modify the data inserted.
Press the "↵ ↵ ↵ ↵ " to confirm date to be entered.
Once the date has been set, it will appear as this example:
DATE
24
TIME
16
After the date, set the time in the same way.
6.1.7
Predicted Values
PREDICTED VALUES
Esc
•
press "↵ ↵ ↵ ↵ " to select PREDICTED VALUES and to choose one of the 3 available authors for the predicted values
calculated for adult subjects:
PREDICTED VALUES
ERS
The symbol
indicates the values currently selected from the 3 available:
KNU : Knudson
ERS : European Respiratory Society
USA : Crapo, Bass, Morris
It is possible to select another author by pressing " "
WARNING
In the case of children or young adults the Knudson mode is always utilised regardless of whether this is the
actual mode selected
6.1.8
Unit Format
UNIT FORMAT
Esc
•
press Ok to select the measurement units required. Select either Kg,cm or Lb,in using the " " key. Press Ok to
confirm.
spirodoc user manual code 980165
Ok
Ok
01 / 04
/ 01 / 04
33 : 15
Ok
KNU
USA
Ok
Rev 4.5
page 22 of 27
. Ok to confirm.
Need help?