ANNEX 1
INFORMATION FOR CORRECT USE IN AN ELECTROMAGNETIC ENVIRONMENT
Guidance and manufacturer's declaration – electromagnetic emissions
The spirodoc is intended for use in the electromagnetic environment specified below.
The customer or the user of the Spirodoc should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
The spirodoc is intended for use in the electromagnetic environment specified below. The customer
or the user of the spirodoc should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
1 kV for input/output
transient/burst
lines
IEC 61000-4-4
Surge
±1 kV differential mode
IEC 61000-4-5
±2 kV common mode
Voltage dips, short
<5 % UT
(>95 % dip in UT)
interruptions and
voltage variations
for 0,5 cycle
on power supply
input lines
40 % UT
(60 % dip in UT)
IEC 61000-4-11
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
NOTE UT is the a.c. mains voltage prior to application of the test level.
spirodoc user manual code 980165
Compliance
Electromagnetic environment – guidance
Group 1
The spirodoc uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
Class B
Not applicable
Not applicable
Guidance and manufacturer's declaration – electromagnetic immunity
Compliance level
±6 kV contact
±8 kV air
Not Applicable
Not Applicable
3 A/m
[3 ] V
[3 ] V/m
Electromagnetic environment –
guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30 %.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
Portable and mobile RF communications equipment should
be used no closer to any part of the spirodoc, including
cables,
than
the
recommended
calculated from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d=[ 3.5 ]
P
3
d=[ 3.5 ]
P 80 MHz to 800 GHz
3
d=[ 7 ]
P 800 MHz to 2,5 GHz
3
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey, should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked
with the following symbol:
Rev 4.5
page 26 of 27
separation
distance