MIR MiniSpir Light User Manual
Hide thumbs Also See for MiniSpir Light:

Advertisement

MiniSpir Light
User Manual Rev. 1.7
Issued on:
09.10.2020
Approved on:
09.10.2020
MINISPIR LIGHT User manual
Rev 1.7
page 1 di 13
User manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MiniSpir Light and is the answer not in the manual?

Questions and answers

Summary of Contents for MIR MiniSpir Light

  • Page 1 MiniSpir Light User manual User Manual Rev. 1.7 Issued on: 09.10.2020 Approved on: 09.10.2020 MINISPIR LIGHT User manual Rev 1.7 page 1 di 13...
  • Page 2 The winspiroLIGHT PC software supplied with the device MUST be installed correctly to the PC before connecting MiniSpir Light to the PC. At the end of the installation, connect the device to the PC and the hardware will be "recognised"...
  • Page 3: Table Of Contents

    Technical features ................................. 9 1.6.1 Features of the spirometer ............................9 1.6.2 Other features ................................. 10 FUNCTIONING OF THE MiniSpir Light ..........................10 Connection to PC ................................ 10 Using the MiniSpir Light .............................. 10 Spirometry Testing ..............................10 Spirometry test interpretation ............................11 DATA TRANSMISSION ..............................
  • Page 4: Introduction

    Important safety warnings MiniSpir Light has been examined by an independent laboratory which has certified the conformity of the device to the Safety Standards IEC 60601-1 and guarantees the EMC Requirements within the limits laid down in the Standard IEC 60601-1-2.
  • Page 5: Danger Of Cross-Contamination

    The correction of the inspired volumes and flows is made automatically as the machine has an internal temperature sensor; the BTPS values are thus calculated. If a 3L syringe is used to make the calibration and if the MiniSpir Light is calibrated correctly then the FVC (syringe) value will be: 3.00 (FVC) x 1.026 (BTPS) = 3.08 L (FVC at BTPS).
  • Page 6: Usb Connection Cable

    Do not use cables that appear to be or are damaged. If a new cable is required, contact your local distributor. Use only cables supplied by MIR, specifically designed to be used with MiniSpir Light. The use of other types of cables can lead to inaccurate measurements.
  • Page 7: Labels And Symbols

    • Environmental procedures as: air conditioning, humidification, conductive floor coverings, non-synthetic clothing • User procedures as: discharging one’s body to a large metal object, using wrist strap connected to earth. MINISPIR LIGHT User manual Rev 1.7 page 7 di 13...
  • Page 8: Product Description

    PRE (before drug administration). The Pre test data relates to percentage variations between the measured results and the predicted values based on the anthropometric data inserted. The POST session is available only on the MiniSpir Light BD version.
  • Page 9: Technical Features

    The personal best values can vary considerably from the predicted values, which are taken from “healthy” subjects. MiniSpir Light is connected to a PC trough a USB port. Data measured by MiniSpir Light are transferred to the PC in real-time.
  • Page 10: Other Features

    Connection to PC WARNING Before connecting MiniSpir Light to a PC, the winspiroLIGHT software must be installed on the PC in order to interface it with the device. To connect the device to the PC, it is sufficient to connect the standard USB socket of the device to one of the USB ports of the When initially making a connection, the PC will, either make an automatic driver installation or request some information.
  • Page 11: Spirometry Test Interpretation

    • Fit the nose clip onto the nose of the subject to ensure that air cannot escape through the nostrils. • Hold MiniSpir Light in one hand as you would a cell phone. The side with the ID label should be in the hand of the user.
  • Page 12: Data Transmission

    All data in the MiniSpir Light is transferred through a USB cable connection. Refer to Paragraph 2.1 of this Manual to connect the device to a PC. The data measured by MiniSpir Light during a spirometry test are sent to the PC in digital form and managed by the winspiroLIGHT software.
  • Page 13: Limited Warranty Conditions

    User manual LIMITED WARRANTY CONDITIONS MiniSpir Light, together with its standard accessories is guaranteed for a period of 12 months if intended for professional use (doctors, hospitals, etc.). The warranty is effective from the date of purchase contained in the relevant sales invoice or proof of purchase.