Troubleshooting Guide - Denver Instrument P-114 Operating Instructions Manual

Pinnacle series analytical and precision balances
Hide thumbs Also See for P-114:
Table of Contents

Advertisement

Available languages

Available languages

System
The System menu lets you configure various features of your balance.
You can assign password protection for menu settings, change the acoustic
signal, restore factory defaults, and adjust the RS-232 interface settings for use
with a computer or printer.
Restoring Factory Defaults
To restore all settings to the factory defaults:
1. Press the Menu key.
2. To open the System menu, press
then press
.
3. To restore the factory settings in the menu, press
"Defaults" is displayed, then press

Troubleshooting Guide

OVER
(CAPACITY ERROR)
UNDER
CAL
ADD WEIGHT
CAL
OUT OF RANGE
Other errors
18
repeatedly until "System" is displayed,
.
The load exceeds the
balance capacity.
Display capacity is
exceeded.
The load is too light.
Calibration/adjustment
started without weight
on pan.
Weight cannot be
detected.
Error has occurred.
repeatedly until
Unload the balance
or remove obstruction.
Decrease weight on
balance.
Check pan position.
Unload the balance and
switch it off and then on
again.
Add weight and press
CAL key again.
Unload pan and add
appropriate weight.
Switch the balance off
and then on again. If the
error persists, contact
a service technician.