Note on Verified Balances Used in
Legal Metrology (PI... A Models):
$ The balance must warm up
for at least 24 hours after initial
connection to AC power.
$ The type-approval certificate for
verification applies to non-auto-
matic weighing instruments only;
for automatic operation with
or without auxiliary measuring
devices, you must comply with the
regulations applicable in your
country.
$ The temperature range indicated
on the verification ID label must
not be exceeded during operation.
Example:
BD BP 200
K 0 to +40°C
Taring the Balance
1. Place a container on the weighing
pan. The balance registers the
weight of the container.
2. Press the TARE key.
3. "TARING" is displayed to indicate
that the balance is being tared.
4. The display shows 0.0000 grams
(or selected weight unit and
resolution) once taring is con-
cluded successfully.
5. When the readout is stable, the
Denver Instrument logo appears in
the upper left corner of the display.
When unstable, a "U" is displayed.
15