Specifications; Tools And Test Equipment Required - OHAUS NV212 Service Manual

Navigator series
Hide thumbs Also See for NV212:
Table of Contents

Advertisement

CHAPTER 1 GETTING STARTED

1.3 TOOLS AND TEST EQUIPMENT REQUIRED

1. Common hand tools are sufficient to disassemble the Navigator scales.
2. RS2323 Interface Kit – PN 83032106 or 71147376

1.4 SPECIFICATIONS

Specifications for the Ohaus Navigator Scales are listed in Table 1-1. When a scale has been
serviced, it must meet the specifications listed in the table. Before servicing the scale, determine
what specifications are not met.
Special Note regarding Approved scales
The specifications for the approved scales below are only for initial testing. These scales must be
tested according to the requirements of the local Weights and Measures authority. Before returning
the scale to service an approved representative of the local Weights and Measures authority must
certify the scale. Ohaus does not generally recommend the repair of an approved scale that
involves replacement of the load cell or PCB.
NV (non-Approved)
Model
Capacity (g)
Repeatability (g)
Linearity (g)
OCL (g)
Zero Drift (g/° C)
Span Sensitivity (g/° C)
NV (CMC Approved)
Model
Capacity
Repeatability (g)
Linearity (g)
OCL (g)
Zero Drift
Span Sensitivity
Ohaus Corporation www.ohaus.com
TABLE 1-1 SPECIFICATIONS
NV212
NV511
210
510
0.02
0.2
±0.02
±0.2
±0.03
±0.3
±0.03
±0.2
±0.01
±0.1
NV212B
NV511B
200
510
0.02
0.2
±0.02
±0.2
1 MPE
1 MPE
1e / 5° C
1e / 5° C
1 MPE
1 MPE
2
NV1101
NV2101
1100
2100
0.2
0.2
±0.2
±0.2
±0.3
±0.3
±0.2
±0.2
±0.1
±0.1
NV1101B
NV2101B
1100
2100
0.2
0.2
±0.2
±0.2
1 MPE
1 MPE
1e / 5° C
1e / 5° C
1 MPE
1 MPE
Navigator Series Service Manual
NV4101
NV5101
4100
5100
0.4
1
±0.4
±1
±0.6
±1.5
±0.4
±1
±0.2
±0.5
NV3101B
NV5100B
3100
5100
0.4
1
±0.4
±1
1 MPE
1 MPE
1e / 5° C
1e / 5° C
1 MPE
1 MPE

Hide quick links: