CHAPTER 4 TESTING
4.5.3 Linearity Test
This test is used to determine the linearity of the unit throughout its operating range. The masses
used to perform this test can be utility masses.
NOTE:
The scale must pass the Precision and Repeatability Tests, and be calibrated before the
Linearity Test may be performed.
Model
NV212
NV311(N or M)
NV511
NV1101
NV2101
NV4101
NV5101
NVL511
NVL1101
NVL2101
NV3100(N or M)
NVL5101
NVL10000
NVL20000
NVT1601
NVT1601(N or M)
NVT3201
NVT3200(N or M)
NVT6400(N or M)
NVT6401
NVT10001
NVT16000
NVT16000(N or M)
NOTE: All masses are nominal values. Be certain to use the same reference mass throughout the
procedure.
2. Place the test mass on the Scale, record the weight and remove.
3. Place Load 1 on the Scale and press TARE.
4. Place the test mass on the Scale, record the weight and remove.
5. Place Load 2 on the Scale and press TARE.
6. Place the test mass on the Scale, record the weight and remove.
7. Place Load 3 on the Scale and press TARE.
8. Place the test mass on the Scale and record the weight.
Ohaus Corporation www.ohaus.com
TABLE 4-4

LINEARITY TEST MASSES

Reference Wt.
50g
75g
125g
250g
500g
1000g
1250g
125g
250g
500g
750g
1250g
2500g
5000g
400g
400g
750g
750g
1500g
1500g
2500g
4000g
4000g
30
Load 1
Load 2
50g
100g
75g
150g
125
250g
250g
500g
500g
1000g
1000g
2000g
1250g
2500g
125
250g
250g
500g
500g
1000g
750g
1500g
1250g
2500g
2500g
5000g
5000g
10000g
400g
800g
400g
800g
750g
1500g
750g
1500g
1500g
3000g
1500g
3000g
2500g
5000g
4000g
8000g
4000g
8000g
Navigator Series Service Manual
Load 3
150g
225g
375g
750g
1500g
3000g
3750g
375g
750g
1500g
2250g
3750g
7500g
15000g
1200g
1200g
2250g
2250g
4500g
4500g
7500g
12000g
12000g