4 TESTING

Before and after servicing a Navigator scale, an operational test and various performance tests should
be made to confirm that the scale meets specifications. Turn the scale on and allow it to warm up for
at least one hour before performing these tests.
NOTE:
Make sure the test area is free from drafts and that the scale rests on a level and vibration-free
surface.
4.1 TEST MASSES REQUIRED
The masses required to test the Ohaus Navigator scales must meet the requirements of ASTM
Class 4 or OIML F2 Tolerance. The mass values are listed in Table 4-1.
NV212
NV311(N or M)
NV511
NV1101
NV2101
NV3100(N or M)
NV4101
NV5101
NVL511
NVL1101
NVL2101
NVL5101
NVL10000
NVL20000
NVT1601
NVT1601(N or M)
NVT3201
NVT3200(N or M)
NVT6400(N or M)
NVT6401
NVT10001
NVT16000
NVT16000(N or M)
Navigator Series Service Manual
TABLE 4-1

CALIBRATION MASS VALUES

Model
Calibration mass
25
Span
Linearity
Calibration masses
200g
100g, 200g
200g
200g, 300g
300g
300g, 500g
500g
500g, 1000g
1kg
1kg, 2kg
2kg
2kg, 3kg
2kg
2kg, 4kg
3kg
3kg, 5kg
300g
300g, 500g
500g
500g, 1000g
1kg
1kg, 2kg
3kg
3kg, 5kg
5kg
5kg, 10kg
10kg
10kg, 20kg
1kg
1kg,1.5kg
1kg
1kg,1.5kg
2kg
2kg, 3kg
2kg
2kg, 3kg
5kg
3kg, 6kg
5kg
3kg, 6kg
5kg
5kg, 10kg
10kg
10kg, 15kg
10kg
10kg, 15kg