B.9 Expanded Resolution (E.pand) - OHAUS NV212 Service Manual

Navigator series
Hide thumbs Also See for NV212:
Table of Contents

Advertisement

APPENDIX B THE SERVICE MENU
There are four settings for the automatic zero setting mechanism (AZSM).
0.5d
The display will maintain a zero indication if the scale does not detect a change of more
than 0.5d. This is the factory default.
1d
The display will maintain a zero indication if the scale does not detect a change of more
than 1d.
3d
The display will maintain a zero indication if the scale does not detect a change of more
than 3d.
OFF
The setting will disable the automatic zero setting mechanism (AZSM). This is generally
used for testing purposes only.
In the Service Menu, press No until a-zero displays.
Press Yes to enter the auto zero setting menu.
The current auto zero setting is displayed.
Press No to increment the setting. Press Yes
to accept the displayed value and advance to
the next menu item, LFT.

B.9 EXPANDED RESOLUTION (E.PAnd)

E.PAnd sets the readability to 1/10
with E.PAnD on it would be .001.) Raw scale data is displayed, zero, tare and other modes/functions
will not operate.
In the Service Menu, press No until E.PAnD appears.
Press Yes to enter the E.PAnD setting menu,
OFF appears. Press No to toggle the setting,
On appears.
Pressing Yes again accepts the On setting.
When the Service Menu is exited the scale will
read with expanded resolution until it is turned
off in the Service Menu.
B.10 RESET
reset returns the scale to its factory default conditions. All the user and service settings will return to
the factory defaults. All reference weights, average piece weights and set points will be reset.
In the Service Menu, press No until reset appears.
Press Yes to enter the reset setting menu, No
appears. Press No to toggle the setting, YES
appears.
To accept a chosen setting (either No or Yes)
press Yes. The display will advance to the
next menu item, End.
Ohaus Corporation www.ohaus.com