C.2 Defining Led Display Version - OHAUS NV212 Service Manual

Navigator series
Hide thumbs Also See for NV212:
Table of Contents

Advertisement

APPENDIX C SOFTWARE SERVICE TOOL INSTRUCTIONS

C.2 DEFINING LED DISPLAY VERSION

If the software version has been updated to version 1.04 or later, the following commands needs to
be sent to the scale to define the LED display version. Please refer to Figure 1-3 for LED version.
Old LED version:
1: MFG
2: (esc)OLED
New LED version:
1: MFG
2: (esc)NLED
C.3 CONFIGURING THE SCALE
If the PC Board has been replaced, the scale needs to be configured. The software tool will send
the PCB the required data for the Model Name entered.
1. When communications are reliable, return to the Replace PCB tab.
2. To configure the scale, enter the model name that appears on the Data Label (e.g. NVL5101,
See figu.C-4) in the Model Name text box (See Fig.C-5).
Ohaus Corporation www.ohaus.com
Enable service mode.
Define current display as old LED display, store this data into EEPROM.
Enable service mode.
Define current display as old LED display, store this data into EEPROM.
Figure C-4. Model Name on Data Label
Figure C-6. Model Name entered