Indicator Lamp

The indicator lamp on the power button has several conditions:
Color
Green
Orange (yellow)
Red
32
State
Description
Static on
Powered on; instrument is calibrated/initialized
and not scanning
Flashing
Scanning (no fault condition exists)
Static on
Instrument initializing (not scanning)
Flashing
Scan in progress while initializing/calibrating
Static on
Initialization or calibration failed. Instrument is still
allowed to operate.
Flashing
Instrument is scanning after initialization or
calibration has failed.