Thermo Scientific Orion Star A113 Instruction Sheet

Thermo Scientific Orion Star A113 Instruction Sheet

Benchtop, portable do meters
Hide thumbs Also See for Orion Star A113:

Advertisement

Quick Links

Display Information

Display Icon
Description
MEAS
Indicates that the meter is in the measurement mode.
SETUP
Indicates that the meter is in setup mode.
CAL
Indicates that the meter is the calibration mode.
MAN
Shown when a manual calibration (Winkler titration) is being done.
AUTO
Shown during water-saturated air calibration. Default setting.
Shows the battery status (more bars = more power remaining).
Blinks when power is low and the battery needs to be changed.
(Batteries included and factory-installed with the portable meter Orion Star A123 DO meter.)
Shown when the meter is running on AC power.
(Adapter included with the benchtop meter Orion Star A113 DO meter.)
Shown when the meter is on AUTO-READ mode. Default setting. AR and unit of measurements
will blink until the reading is stable. When the reading is stable it is held on the screen and AR is
lit. Press
to take a new reading.
Unit of measurement will blink until the reading is stable.
READY
When the reading is stable, READY is lit.
Appears during and after a calibration has been performed.
Displayed when a reading is stored into the memory.
Displayed when viewing stored readings.
Secondary display
Shows temperature reading in measurement mode and setup menu in setup mode.
Primary display
Measured value in selected mode.
Instructions
Located below the primary display. These phrases aid in the setup menu and calibration modes.
©2011 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. & its subsidiaries.
North America
Netherlands
Water Analysis Instruments
166 Cummings Center
Tel: (31) 033-2463887
www.thermoscientific.com/water
Beverly, MA 01915 USA
info.water.uk@thermo.com
Toll Free: 1-800-225-1480
China
Tel: 1-978-232-6000
Tel: (86) 21-68654588
info.water@thermo.com
9 0 0 1 : 2 0 0 8
wai.asia@thermofisher.com
IS-267280-001 Rev. B 06-11
India
Japan
Tel: (91) 22-4175-8800
Tel: (81) 045-453-9175
wai.asia@thermofisher.com
wai.asia@thermofisher.com
Singapore
Australia
Tel: (65) 6778-6876
Tel: (613) 9757-4300
wai.asia@thermofisher.com
in Australia (1300) 735-296
InfoWaterAU@thermofisher.com
Thermo Scientific Orion Star A113
English
Benchtop & Star A123 Portable DO Meters
DO
Instruction Sheet

Preparation

1. Power source:
a. Power adapter (included with Orion Star A113 benchtop DO meters,
sold separately for Orion Star A123 portable DO meters) – Select the
appropriate wall socket plug. Slide off the clear plastic cover, and slide
on the plug plate into the groove on the back of the adapter.
b. Batteries (factory installed on Orion Star A123 portable DO meters,
sold separately for Orion Star A113 benchtop DO meters) – Select
four AA batteries. Confirm that the meter is off and remove the
battery compartment cover. Insert batteries as shown in the battery
compartment housing.
2. Prepare the DO probe according to the probe instructions. In general:
a. Unscrew the membrane cap from the DO probe, fill the membrane
cap about ¾ full with polarographic electrolyte solution and screw the
membrane cap onto the DO probe.
b. Connect the DO probe to the meter and allow the probe to polarize for
about 30 to 60 minutes.
A DO reading of zero and no change in temperature indicates that the
DO probe is not fully connected to the meter. Disconnect the DO probe and then
firmly reconnect the DO probe to the meter.
3. Prepare the calibration sleeve or BOD bottle.
a. For the calibration sleeve, remove the cap from the sleeve and remove the
sponge from the cap. Wet the sponge with distilled water and squeeze out
excess water. Reassemble the calibration sleeve and insert the DO probe.
b. For a BOD bottle, fill the bottle with about 50 mL of distilled water. Insert
the DO probe into the BOD bottle. Use a Thermo Scientific Orion BOD bottle
adapter if the DO probe does not fit directly into the bottle. Make sure that
the probe is suspended about half an inch above the distilled water and
there is no water on the surface of the DO probe membrane.
4. Connect the appropriate item as labeled on the meter and as shown in Figure 1.

Overview

1. To turn on or off the meter, press
.
2. To exit any meter function, press
to return to the measurement mode.
3. The meter mode is shown at the top of the meter:
MEAS – for measurement mode
SETUP – for setup mode
CAL – for calibration mode
4. The meter can be calibrated using water-saturated air or using the Winkler titration
method for a manual calibration.
5. Press
to switch between reading in % saturation, mg/L, solution
temperature or membrane temperature. ("m" will appear in front of the
membrane temperature reading.)
ation
Figure 1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Orion Star A113 and is the answer not in the manual?

Questions and answers

Summary of Contents for Thermo Scientific Orion Star A113

  • Page 1 Instruction Sheet Preparation 1. Power source: a. Power adapter (included with Orion Star A113 benchtop DO meters, sold separately for Orion Star A123 portable DO meters) – Select the appropriate wall socket plug. Slide off the clear plastic cover, and slide...
  • Page 2 Orion Star A113 Benchtop and Orion Star A123 Portable DO Meters English Entering Barometric Pressure Memory Feature DO Calibration Selection Dissolved oxygen readings are dependent on barometric pressure. This needs to be entered This meter stores up to 50 readings.

This manual is also suitable for:

Star a113

Table of Contents