Disabling Audible Alarms

http://www.mallinckrodt.com/respiratory/resp/Serv_SuppProductManuals.htm
N-595
1. Press the ALARM SILENCE button until the alarm silence
duration setting is displayed. Alarm silence durations that are
available are OFF, 30, 60, 90, and 120 seconds.
2. Press and hold the ALARM SILENCE button and the ADJUST
UP button to increase the alarm silence duration setting.
3. Press and hold the ALARM SILENCE button and the ADJUST
DOWN button to decrease the alarm silence duration setting.
Note: Releasing the ADJUST UP or ADJUST DOWN button sets the
alarm silence duration.
Setting the alarm silence duration to OFF means that the monitor will
produce no audible alarms.
Note: The ability to set the alarm silence duration to OFF can be
enabled or disabled by qualified service personnel as described
in the service manual. The current copy of the service manual
is available on the Internet at: