5
Vial Handling

Vial Size Errors

62
The HS checks vial height:
• When loading a vial onto the sampling probe (12 vial
model)
• When grasping a vial with the gripper (111 vial model)
If the vial height does not match the expected value for the
vial size defined in the method, a vial size error is logged.
The HS proceeds in accordance with the defined sequence
action for vial size.
Note that 20- mL and 22- mL vials are the same height. The
HS cannot distinguish between them. (An internal capacity
of 20 mL or 22 mL is within the range of exterior
dimensions for either vial type.)
Agilent 7697A Troubleshooting