Maintenance Log Entries - Agilent Technologies 7697A Troubleshooting Manual

Headspace sampler
Table of Contents

Advertisement

3
Log Entries and Errors

Maintenance Log Entries

Table 3
Event log entries and descriptions
Message
<Component> serviced
Firmware updated: x.xx.xx
<Component> due, or
<Component> service due
Tray calibrated
Tray calibration failed
Leak Test Passed
Leak Test (Part 2) Passed
Leak Test failed
##
30
The maintenance log records maintenance operations
performed and logged by the user. The log contains up to
128 records. If at capacity, any new entries overwrite the
oldest ones. The log remains until specifically cleared.
The HS always logs messages such as leak test results and
firmware updates. However, "serviced" and "service due"
messages are logged only when using the Early Maintenance
Feedback (EMF) features of the instrument. (See the
Advanced Operation
counter, the HS assumes that maintenance has been
performed.
Description
Indicates that the counter was reset for the indicated component.
Indicates an the instrument firmware update to the revision shown.
The EMF counter for the indicated component has been reached.
Indicates that a tray calibration was successfully performed.
Indicates that a tray calibration failed.
Indicates a leak test step failed, where ## is the part of the test which failed.
guide.) When the user resets a resource
Agilent 7697A Troubleshooting

Hide quick links: