Retention Times Not Repeatable

Agilent 7697A Troubleshooting
• Check the GC. See the GC's troubleshooting information.
GC problems can include:
Leaks in the inlet, including septum
Leaks around the inlet liner, if applicable
Gas supply pressure
Wrong liner type used, if applicable
First run effects (has the GC stabilized?)
• Check the HS sequence log, event log, and status displays
for messages. Especially check the sequence log for
method deviations and dynamic leak check failures.
• If using HS carrier gas control, confirm that the current
HS column configuration matches the column actually
installed in the GC.
• Check for leaks in the transfer line to the GC inlet,
including the connection to the inlet.
• Check for leaks in the other parts of the sampling system
(probe, six port valve, and lines and fittings between).
• Enable dynamic leak checking. This feature verifies that
each sample vial is leak- free.
If not enabled, check the sample vial for cracks or
other leaks. See the
If using dynamic leak checking, make sure the entered
leak rate is appropriate for the sample and its
preparation technique.
Chromatographic Symptoms
Operation
Guide.