Retention Times Not Repeatable
Agilent 7697A Troubleshooting
• Check the GC. See the GC's troubleshooting information.
GC problems can include:
•
Leaks in the inlet, including septum
•
Leaks around the inlet liner, if applicable
•
Gas supply pressure
•
Wrong liner type used, if applicable
•
First run effects (has the GC stabilized?)
• Check the HS sequence log, event log, and status displays
for messages. Especially check the sequence log for
method deviations and dynamic leak check failures.
• If using HS carrier gas control, confirm that the current
HS column configuration matches the column actually
installed in the GC.
• Check for leaks in the transfer line to the GC inlet,
including the connection to the inlet.
• Check for leaks in the other parts of the sampling system
(probe, six port valve, and lines and fittings between).
• Enable dynamic leak checking. This feature verifies that
each sample vial is leak- free.
•
If not enabled, check the sample vial for cracks or
other leaks. See the
•
If using dynamic leak checking, make sure the entered
leak rate is appropriate for the sample and its
preparation technique.
Chromatographic Symptoms
Operation
Guide.