ABB 6X1EH Operating Instructions Manual page 22

600t en series hart pressure transmitters
Hide thumbs Also See for 6X1EH:
Table of Contents

Advertisement

ADDENDUM FOR COMETER - ANALOG LCD INDICATOR WITH HART PROGRAMMING CAPABILITY
Under ConF PERCENT option, the user can decide for linear
or SQR output. When SQR output is selected, the output is
linear from 0 to 20% (to 4% of input).
Refer to Fig. 6 - ConF METER for details on the procedures.
ConF XMTR - TRANSMITTER CONFIGURATION
Four are the operations under the ConF XMTR menu:
CONF, TRIM, REVIEW and PV.
By pressing ENTER on the ConF XMTR menu, the string
LOADING appears on the display, with the blinking star (*)
indicating communication activity, i.e. the CoMeter is reading
the transmitter information.
See below a list of the available operation under the selected option:
CONF menu
Change LRV
Change URV
Change DAMPING
Change UNITS
Change OUTPUT
Use PREV or NEXT key to scroll through the options and ENTER key to change or view the values.
The procedure to change the numeric value remains the one already explained for PASSWORD operation, i.e., the cursor starts
blinking on the most significant digit, then use the NEXT and PREV key to increase or decrease the value of the single digit (the
minus sign(-)automatically appears or disappears when the value increases above 9 or decreases below 0, as well as for the
decimal point(.). Use the ENTER key to move the cursor to the next digit, use the ESC key to move back to the previous digit.
An ENTER on the last digit will cause the value to be sent to the transmitter.
Refer to figures 7, 8, 9 and 10 for details.
CONF
ENTER
ESC
LOOP
IN_MAN.
ENTER
ESC
NEXT
NEXT PREV
0.000
PREV
LRV
ENTER
ESC
00.000
01.000
01.000
01.000
01.000
LRV
ENTER
NEXT PREV
TO MODIFY DIGIT,
DECIMAL POINT,
MINUS SIGN
- 22 -
TRIM menu
Reranging (RERANG.)
Loop test (LOOPTST)
Output trim (OUTTRIM)
Zero adjustment (SNSZERO)
NEXT PREV
20.000
URV
ENTER
ESC
20.000
25.000
25.000
25.000
25.000
URV
ENTER
NEXT PREV
TO MODIFY DIGIT,
DECIMAL POINT,
MINUS SIGN
Then the CONF option appears.
Using PREV or NEXT key, the user can select CONF, TRIM,
REVIEW or PV option, and with the ENTER key he moves into
the menu.
When entering CONF and TRIM menu a message "LOOP
IN_MAN" appears to remind that a modification can change
the transmitter output, so for security the loop should be put in
Manual.
REVIEW menu
TAG 8
Final Assembly Nr. (XMTR N.)
Sensor Serial Nr. (SENS N.)
Up/Down scale (UP/DOWN)
UNITS
LRV
URV
LRL (See Sensor Units)
URL (See Sensor Units)
DAMPING
OUTPUT
NEXT PREV
0.0000
DAMPING
ENTER
ESC
0.0000
0.2000
0.2000
0.2000
0.2000
DMP SEC
ENTER
NEXT PREV
TO MODIFY DIGIT,
DECIMAL POINT,
MINUS SIGN
Fig. 7 - CONF menu
PV menu
Primary variable (PRIMARY)
Secondary variable (2ND)
Tertiary variable (3RD)
Fourth variable (4TH)
NEXT PREV
UNITS
ESC
ENTER
ESC
KPA
TORR
ATM
MPA
IN H2O
KG/CM2
ENTER
NEXT
PREV
OUTPUT
ENTER
LINEAR
SQR

Advertisement

Table of Contents
loading

This manual is also suitable for:

6x1ed6x1ee6x1eg6x1es622ed6x1ea

Table of Contents