Operating Instructions Safety Cabinet | 1300 Series B2
4.
Performance Features and Safety Precautions
When surface disinfecting the Biosafety Cabinet:
Avoid splashing the disinfecting solution on skin or clothing.
Ensure adequate ventilation.
Carefully follow the disinfectant's safety instructions.
Always dispose of disinfecting solutions in accordance with local and
national laws.
DO NOT allow disinfectants with high concentrations of free chlorine to
contact the stainless steel components of the Biosafety Cabinet for a
long period of time. Free chlorine will corrode stainless steel after
extended contact.
The electrical receptacle cover may be removed to facilitate surface
decontamination.
Biosafety Cabinets should be decontaminated for any of the following reasons:
Before maintenance work requiring entry into contaminated areas.
Before exhaust HEPA filter changes.
Before performing certification tests requiring entry into contaminated
areas.
Before relocating the cabinet.
Before changing research programs.
After the gross spill of biohazardous material or toxic chemicals.