Protective Equipment; Proper Use - Pfeiffer HiPace 300 Operating Instructions Manual

Turbopump
Table of Contents

Advertisement

Safety

2.2 Protective equipment

kg

2.3 Proper use

6
• Keep leads and cables well away from hot surfaces (> 70 °C).
• Never fill or operate turbopump with cleaning agent.
• Do not operate the turbopump with open high vacuum flange.
• Do not carry out any unauthorized modifications or conversions to the pump.
• When returning the turbopump observe the shipping instructions.
Determined situations concerning the handling of vacuum pumps require wearing
of personal protective equipment. The owner, respectively the employer are obli-
gated to provide adequate equipment to any operating persons.
Danger to health by hazardous substances during maintenance or installation
Depending on the process vacuum pumps, components or operating fluids can be con-
taminated by toxic, reatcive or radioactive substances.
 Wear adequate protective equipment during maintenance and repairs or in case of
reinstallation.
Risk of injury through falling objects
When transporting vacuum pumps by hand, there is a danger through loads slipping and
falling down.
 Carry small and mid-size vacuum pumps two-handed.
 Carry vacuum pumps > 20 kg by a suitable lifting device.
 Wear safety shoes with steel toe cap according to directive EN 347.
Risk of injury through hot surfaces
Vacuum pumps can become hot during operation.
 Allow the pump to cool before maintenance and repairs.
 If necessary wear protective gloves according to directive EN 420.
Risk of injury through sharp edges
Rotor and stator disks of turbopumps have very sharp edges.
 Before any working wait for the complete standstill of the pump.
 Do not reach in the high vacuum flange.
 If necessary wear protective gloves according directive EN 420.
CE conformity
The manufacturer's declaration of conformity becomes invalid if the operator modifies
the original product or installs additional components.
 Following installation into a plant and before commissioning, the operator must
check the entire system for compliance with the valid EU directives and reassess it
accordingly.
• The vacuum pump may only be used to generate a vacuum.
• Only operate the turbopump with an approved backing pump.
DANGER
WARNING
CAUTION
CAUTION
NOTE

Advertisement

Table of Contents
loading

Table of Contents