Modes of Measurement

Mode
(Frequency)
1
Dial & Electronic
2
Impulse Tool
4
5
6
Click & Cam
3
TIP:
To simplify operation DISABLE all modes of measurement that are not required.
See SETTINGS, MODES ENABLE/DISABLE menu.
TIP:
The peak reading can be set up to automatically reset by changing PEAK MEMORY RESET
from MANUAL to AUTO. See SETTINGS, THRESHOLDS
TIP:
For slower operation of any AUTO RESET mode, change AUTO RESET HOLD TIME to 4
SECOND. See SETTINGS, THRESHOLDS menu.
TIP:
In CLICK & CAM mode the serial port will only output for a genuine first peak.
Pressing enter or ZERO will not send an output.
TIP:
If torque wrench readings are inconsistent in CLICK & CAM mode, change FIRST PEAK
SENSITIVITY in the SETTINGS, THRESHOLDS menu to be less sensitive i.e. MEDIUM or LOW.
This will compensate for torque wrench sensitivity.
Mode
TRACK
(500 Hz)
(500 Hz)
Holds the highest reading
until RESET by the user.
[The highest reading can
be automatically reset if
(500 Hz)
AUTO is selected for
PEAK MEMORY RESET.
After the value returns to
Clutch Tool
zero, the memorised
(500 Hz)
reading is held for the
AUTO RESET HOLD
TIME, then resets].
Stall Tool
(500 Hz)
Hold 1
'AUTO RESET HOLD
(500 Hz)
TIME', then resets.
How it Works
Follows signal.
st
peak for the
Visual Representation
AUTO RESET HOLD TIME