TTT SERIES 3 OPERATORS HANDBOOK

MODES OF MEASUREMENT:- __________________________________________________

Mode
Mode
(Frequency)
TRACK
(500 Hz)
1
Dial &
Electronic
2
(500 Hz)
Impulse Tool
(500 Hz)
4
Clutch Tool
(500 Hz)
5
Stall Tool
(500 Hz)
6
Click & Cam
(500 Hz)
3 3 3
TIP. To simplify operation DISABLE all modes of measurement that are not required.
See SETTINGS, MODES ENABLE/DISABLE menu.
TIP. The peak reading can be set up to automatically reset by changing PEAK MEMORY
RESET from MANUAL to AUTO. See SETTINGS, THRESHOLDS
TIP. For slower operation of any AUTO RESET mode, change AUTO RESET HOLD TIME
to 4 SECOND. See SETTINGS, THRESHOLDS menu.
TIP. In CLICK & CAM mode the serial port will only output for a genuine first peak.
Pressing enter or ZERO will not send an output.
TIP. If torque wrench readings are inconsistent in CLICK & CAM mode, change FIRST
PEAK SENSITIVITY in the SETTINGS, THRESHOLDS menu to be less sensitive i.e.
MEDIUM or LOW. This will compensate for torque wrench sensitivity.
How it works.
Follows signal.
Holds the highest
reading until RESET by
the user. [ The highest
reading can be
automatically reset if
AUTO is selected for
PEAK MEMORY
RESET. After the value
returns to zero, the
memorised reading is
held for the AUTO
RESET HOLD TIME,
then resets ].
st
Hold 1
peak for the
'AUTO RESET HOLD
TIME', then resets.
PAGE 14 OF 28
Issue 1
JANUARY 2006
Visual representation
AUTO RESET HOLD