EXCELL PRECISION CO., LTD
2-4-3 Keyboard buzzer type
Press
TARE
Using
or
ON/ZERO
To enter the setting
Press
TARE
Press"00"
Press
key
TARE
to complete the setting
PW Service Manual
UNITS
Keyboard buzzer type fitted
Display shows current type
Keyboard buzzer type fitted
Use
or
ON/ZERO
0 ⇒ Pulse signal (DC type, PT-134OP)
1 ⇒ Contiunous signal (AC type, UIC-
4201)
4 Note the type selected should match
that which is fitted to the hardware in
the indicator
ON/ZERO key ⇒ Step Up key
(Selects number 0,1,2,3,4,5,6,7,8,9)
UNITS key ⇒ Step Down key
(Selects number 0,1,2,3,4,5,6,7,8,9)
TARE key ⇒ Right key, flashing digit move to the right
NET/GROSS key ⇒ Left key, flashing digit move to the left
29
to alter value
UNITS
SME300000016