2-1-3 Modify "Customized Weight Units" Setting

Press
key
TARE
Refer to "Note 4"
Press
key
TARE
Refer to "Note 5"
Press
key
TARE
Refer to "Note 5"
Press
key
TARE
Refer to "Note 6"
Press
key
TARE
Users can either exit
the setting or
continue to modify
it
PW Service Manual
.
6
6
6
6
6
6
.
.
The display shows the maximum
capacity of the scale + minimum
division set previously for 2
lb
seconds.
The display shows the number of weight
units and the weight unit for calibration set
previously
For example:
5 ⇒ 5 weight units
2 ⇒ set "lb" as weight unit for calibration
Show the code of weight unit for calibration
for 2 sec
For example:
019200 ⇒ 192.00 oz
Enter the mass value
12 lb = 192 oz
019200 ⇒ 192.00 oz
For example:
5 ⇒ minimum division
2 ⇒ 0._0.00
4 ⇒ " lb "
0 ⇒ full range
3 ⇒ save at the third position
The display shows the maximum
capacity of the scale + minimum
division set previously for 2 seconds.,
then back to upper level.
lb
ON/ZERO key ⇒ Increment number by 1
UNITS key ⇒ Decrement number by 1
TARE key ⇒ Move cursor one place to the right
NET/GROSS key ⇒ Move cursor one place to the left
19
⇒ " lb "
⇒ " g "
⇒ " kg "