Excell PW Service Manual page 26

Ip67 waterproof
Table of Contents

Advertisement

EXCELL PRECISION CO., LTD
The weight factor is arranged as follows:-
1 = The weights are equal.
2 = The next weight is twice as big as the first weight
3 = 3 times as big as the first weight
4 = 4 times...
5 = 5 times ...
6 = 6 times..
7 = 7 times...
8 = 8 times...
9 = 9 times....
A = 10 times...
B = 11 times...
C = 12 times...
D = 13 times...
E = 14 times...
F = 15 times as big as the first weight
Example: 30kg scale to be linearised with the weight values shown in brackets:-
Ex1: 30kg (10kg ,10kg ,10kg)
Display
Key Press
03 CLn
TARE
L0
TARE
L1 1
TARE
L2 1
TARE
L3 1
TARE
L4 1
NET/GROSS
03 CLn
Ex2: 30kg (5kg, 10kg ,10kg, 5kg)
Display
Key Press
03 CLn
TARE
L0
TARE
L1 1
TARE
L2 2
TARE
L3 2
TARE
L4 1
TARE
L5 1
NET/GROSS
03 CLn
Ex3: 30kg (5kg, 5kg ,10kg, 10kg)
Display
Key Press
03 CLn
TARE
L0
TARE
L1 1
TARE
L2 1
TARE
L3 2
TARE
L4 2
TARE
L5 1
NET/GROSS
03 CLn
PW Service Manual
Note
Into linearity calibration
First point (zero), remove weigh pan and press the Tare key
Put 10kg on and press the Tare key
Put 10kg on and press the Tare key
Put 10kg on and press the Tare key
To finish linearity adjustment (4 points linearity calibration)
Note
Into linearity calibration
First point (zero), remove weigh pan and press the Tare key
Put 5kg on and press the Tare key
Put 10kg on and press the Tare key, 2 is the rate of L1
(10kg is 2 x 5kg, which was used in L1)
Put 10kg on and press the Tare key
Put 5kg on and press the Tare key
To finish linearity adjustment (5 points linearity calibration)
Note
Into linearity calibration
First point (zero), remove weigh pan and press the Tare key
Put 5kg on and press the Tare key
Put 5kg on and press the Tare key
Put 10kg on and press the Tare key
Put 10kg on and press the Tare key
To finish linearity adjustment (5 points linearity calibration)
24
SME300000016

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents