Approval Configuration - Excell PW Service Manual

Ip67 waterproof
Table of Contents

Advertisement

EXCELL PRECISION CO., LTD

2-4-2 Approval configuration

Press
key
TARE
Using
ON/ZERO
to input a parameter
Press
key
TARE
Input the parameter "00"
Press
key
TARE
to complete the setting
Brazil approved model: In the counting mode, the indicator is not able to calculate the count
n
value if the unit weight is less than 0.1e.
PW Service Manual
or
UNITS
Approval configuration
Display shows current setting
Approval configuration
Using
ON/ZERO
0 ⇒ Non-approved model
1 ⇒ OIML or NTEP approved model,
2 ⇒ Sri Lanka approved model, zero can
be accepted within a range of the zero
calibrated point +/- 3%.
3 ⇒ Sri Lanka approved model, zero can
be accepted within a range of the zero
calibrated point +/- 4%.
4 ⇒ Brazil approved model.
5 ⇒ Approved model. In the tare mode, the
gross weight will be displayed after
pressing the
seconds, the net value will be
displayed automatically.
External weight calibration is not available
n
for all approved models.
ON/ZERO key ⇒ Increment number by 1
UNITS key ⇒ Decrement number by 1
TARE key ⇒ Move cursor one place to the right
NET/GROSS key ⇒ Move cursor one place to the left
28
or
to input the setting
UNITS
key. After 5
NET/GROSS
SME300000016