EXCELL PRECISION CO., LTD

2-1-1 Weight Units Setting

There is no resolution limitation, when the weight units "kg, g, lb, lb/oz" are selected.
The weight units "oz, GN, dwt, and ct" are only available on indicators with less 1/10,000
external resolution.
Press
key
TARE
Refer to "Note 1"
Press
key
TARE
Refer to "Note 2"
Press
key
TARE
Refer to "Note 3"
Press
key
TARE
Input the parameter "00
Press
key to
TARE
complete the setting.
PW Service Manual
kg
.
.
12
The display shows the maximum
capacity of the scale + minor
division size and the weight units
for 2 seconds.
Example:
Two weight units "kg" & "lb", with
"kg" as the units used for weight
calibration.
Example:
015000 ⇒ 15.000 kg
For example:
1 ⇒ minimum division
3 ⇒ decimal point position
2 ⇒ range setting
The display shows the maximum
capacity of the scale + minimum
division and the unit weight for 2
seconds.
ON/ZERO key ⇒ Increment number by 1
UNITS key ⇒ Decrement number by 1
TARE key ⇒ Move cursor one place to the right
NET/GROSS key ⇒ Move cursor one place to the left